1
ENGLISH
Elipar
™
DeepCure-L
LED Curing Light
en
Table of Contents
Page
Safety Instructions
1
Glossary of Symbols
2
Product Description
3
Fields of Application
3
Technical Data
3
Charger
3
Handpiece
3
Charger and Handpiece
4
Transport and Storage Conditions
4
Installation of the Unit
4
Factory Settings
4
Initial Steps
4
Charger
4
Light Guide / Handpiece
4
Battery Charging
4
Battery Power Level Display on Handpiece
5
Operation
5
Selection of Exposure Time
5
Activating and Deactivating the Light
6
Inserting and Removing the Light Guide
from/into the Handpiece
6
Positioning the Light Guide
6
Testing of Light Intensity
6
Sleep Mode
7
Audible Signals - Handpiece
7
Troubleshooting
7
Maintenance and Care
8
Care of the Handpiece
8
Cleaning the Light Guide
8
Clean Handpiece and Glare Shield
9
Storage of the Handpiece during
Extended Periods of Non-Use
9
Return of Old Electric and
Electronic Equipment for Disposal
9
Customer Information
10
Warranty
10
Limitation of Liability
10
Safety Instructions
PLEASE NOTE!
Prior to installation and start-up of the device, please
read these instructions carefully.
As with all technical devices, the proper function and
safe operation of this device depend on the user’s
compliance with the standard safety procedures as well
as the specific safety recommendations presented in
these Operating Instructions.
1. Use of the device is restricted to trained personnel in
accordance with the instructions below. The manu-
facturer assumes no liability for any damage arising
from any other or improper use of this device.
2. The charger must be accessible at all times. Do not
use the charger for any use other than the charging
of the Elipar
™
DeepCure-L handpiece. Disconnect the
handpiece from the mains by unplugging the charger
from the electrical outlet.
Treating patients using the handpiece while it is still
connected to the charger is prohibited for safety
reasons. Light-curing is possible only if the charger
has been disconnected.
3. Use only the charger (AC adapter plug) which is
provided with the device. The use of any other
charger can result in damage to the battery.
4. CAUTION. Do not stare at source. May be harmful
to the eyes. Restrict exposure to the area of the
oral cavity in which clinical treatment is intended.
Protect patient and user from reflection and intensive
scattered light by taking the appropriate precautions,
e.g., glare shields, goggles, or coverings.
5. CAUTION! As is the case for all high-intensity light-
curing devices, the high light intensity is accompanied
by heat generation on the exposed surface. This heat
can result in irreversible damage if there is longer
exposure in the proximity of the pulp or soft tissue.
The exposure times given in the manufacturer’s
instructions must be observed exactly to avoid any
such damage.
Uninterrupted exposure times of the same tooth
surface in excess of 20 seconds and direct contact
with oral mucosa or skin must be strictly avoided.
Scientists working in this field are in agreement
that the irritation caused by heat generated during
light curing can be minimized by taking two simple
precautions:
• Polymerization with external cooling from an air flow
• Polymerization at intermittent intervals
(e.g., 2 exposures lasting 10 seconds each instead
of 1 exposure lasting 20 seconds).
6. Elipar DeepCure-L may be operated only with the
supplied light guide or original 3M ESPE Elipar
DeepCure-L replacement and accessory light guide.
ENGLISH
en
Содержание Elipar DeepCure-L
Страница 77: ...76 reset 2 LED LED...
Страница 79: ...78 START Protemp Crown RelyX Unicem 3M ESPE...
Страница 80: ...79 el Elipar DeepCure S Sleep Mode sleep mode 5 sleep mode START 1 5 2 10 3 15 10 START 2 2 TIME START...
Страница 81: ...80 2 5 10 3M ESPE START reset...
Страница 82: ...81 el Elipar DeepCure L START 2 2 3 START 3 START Elipar DeepCure L...
Страница 83: ...82 pH 5 5 pH 8 5 134 C 273 F 500 2 10 G G Cidezyme Enzol Johnson Johnson 10 G Cidex OPA Johnson Johnson 10...
Страница 84: ...83 el 3M Deutschland GmbH C 134 C 273 F 20 121 C 250 F 3 132 C 270 F 134 C 273 F 1 C...