PANNORAMIC® DESK DW II 3.0.4
SCANNING MODES AND SETTINGS
MARCH 30, 2021
59
•
The
Use saved compensation image
function enables the use of previously recorded
reference images for all slides. Use this function when the slide is completely covered by
the specimen (for example, blood smear) thereby making compensation image capture
impossible. For such slides, use an empty slide with coverslip to grab a compensation
image. If you select this, the software will not create new compensation images for each
slide. Compensation images are reference images used for white balancing and shading.
•
With
Save focus quality
the measurement of focus quality (correctness of average focus
on layers) is saved. This option is inactive when
Extended focus
option is activated.
•
Save image in original Bayer size
function is non applicable for DESK.
5.1.2
Microscope control panel
The Microscope Control panel includes the main control features of the microscope.
5.1.2.1
Preview window
The Preview panel
offers options for threshold setting and selection of the area to
be scanned.
Figure 43 – Preview panel