PANNORAMIC® DESK DW II 3.0.4
TROUBLESHOOTING AND MAINTENANCE
MARCH 30, 2021
102
Warning!
Do not modify the camera driver settings, as the default factory settings give the most
accurate images.
Virtual slides show poor stitching in Pannoramic® Viewer.
•
Scan camera not properly adjusted (for example, due to loose fixture)
•
Contact 3DHISTECH Service for support.
Slide is often positioned incorrectly.
•
Guiding fault in transport system
•
Contact 3DHISTECH Service for support.
Digitization process is extremely slow.
•
Extended focusing
option is activated.
•
Deactivate the
Extended focusing
option if not required.
•
Z-stack
option is activated.
•
Deactivate the
Z-stack
option if not required.
Image is out of focus.
•
Focus range limit is activated.
•
Deactivate the
Limiting focus range
option.
•
Focus range setting is incorrect
•
Change focus range setting.
All individual images show a shading effect.
•
Algorithm failed to find empty image fields
•
Capture a reference image with a blank slide.