![3DHISTECH PANNORAMIC DESK DW II 3.0.4 Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/3dhistech/pannoramic-desk-dw-ii-3-0-4/pannoramic-desk-dw-ii-3-0-4_user-manual_3121887101.webp)
PANNORAMIC® DESK DW II 3.0.4
TROUBLESHOOTING AND MAINTENANCE
MARCH 30, 2021
100
The product is not equipped with protection against hazardous materials (infectious, toxic,
radioactive, corrosive substances or other effects damaging to health). Decontaminate
infectious materials after use.
Slides that are not used any longer do not classify as domestic waste and must be treated
as hazardous materials and disposed in compliance with the currently valid legal
requirements.
Dust, specks, and humidity may decrease or harm the operating capability of the device.
Protect it with dust cover when not in use. Switch the power off before covering it.
Avoid touching the lamp bulb. Wait for about 15 minutes to allow the bulb to cool down and
disconnect the power plug before changing the bulbs.
Defective devices do not classify as domestic waste. Dispose of the device in compliance
with the currently valid legal requirements.
8.2.1
Cleaning the device
Before cleaning, switch off the device and pull the plug out.
Do not let liquid penetrate into the system.
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
To clean greasy surfaces, it is recommended to apply
soapy water
on a piece of lint-free
cotton cloth and gently rub the surface.
8.2.2
Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage
temperature above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.