Pannoramic SCAN II 2.0 User’s Guide
5 Scanning modes and settings
•
TMA Grid Setting
:
This function is enabled only if a TMA grid has already been placed
on the preview image. Set grid parameters (row and column numbers) and spot shape (round
or rectangle).
Figure 28 – TMA tool: Grid selector
Delete TMA grid by clicking the
X
at the top right corner, and rotate grid by clicking the icon
at the bottom right corner. Enter the number of rows and columns in the
Rows
and
Columns
fields or select the appropriate values with the arrows.
Enter the distance between two spots on the X and the Y axis in µm in the
Core distance
fields or select the appropriate values with the arrows. Enter the average diameter of the
spots in the
Core size
field or select the appropriate value with the arrows.
•
Clear scanmap
:
The entire scanmap area will be cleared and reset to default.
•
Use selection box restriction
: The preview image shows the area that will be digitized
with a red and/or an orange mask. If the
Use selection box
is enabled, and the crop function
is selected, a black frame displayed in the image, and only the area inside the box is scanned.
It is useful if you only want to scan the area that is equivalent with the size of the coverslip, so
external specks will not be scanned.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
60(142)
For research and education uses only, not for use in diagnostic procedures.