Pannoramic SCAN II 2.0 User’s Guide
5 Scanning modes and settings
◦
Brightfield ICC Profile
Select an applicable preset color profile to be used for Brightfield scanning from the drop-
down menu. Settings according to the profile has effect on the color set, color
temperature, and gamma value of the scanned image.
Preset profiles
are the following:
▪
Calibrated Linear
: Color scheme is defined by calibration (Gamma value is not
included)
▪
sRGB standard
: Color scheme is defined by calibration (standard Gamma value -2.2-
is applied)
▪
sRGB standard Microscope Like
: Color scheme is defined by calibration (standard
Gamma value / White Point: D50)
▪
Legacy
: a Color scheme calibrated for the previous software version is applied
▪
Custom
: parameters are freely customizable as follows:
◦
Color Scheme
▪
Calibrated
: Scheme is defined by the color calibration slide
▪
Legacy
: a Color scheme calibrated for the previous software version is applied
▪
Raw
: the color information of the recorded raw image is applied
◦
White Point
▪
D65
: color temperature according to a standard illuminant D65 white point value is
applied (6500K / daylight)
▪
D50
: color temperature according to a standard illuminant D50 white point value is
applied (5000K / horizontal daylight)
◦
Gamma
▪
Standard
: a value of 2.2 is used
▪
Medium
: a value of 1.8 is used
▪
Off
: a value of 1.0 is used
Note:
If Lumencor SPECTRA 6 is configured as illumination unit, the
corresponding
Enable Lumencor hardware trigger
option is displayed
on the
Filter settings
form.
June 4, 2018 - Rev. 2
3DHISTECH Ltd.
106(142)
For research and education uses only, not for use in diagnostic procedures.