Pannoramic SCAN 150 1.15 User’s Guide
5 Troubleshooting and Maintenance
Caution!
If Fluorescence scanning option is available, create new compensation
images after replacing the bulb in HXP 120. For more information see
section
4.2.2
Image Compensation / Fluorescent Compensation
.
5.3.4 Removing a slide
If a slide got jammed or has suffered breakage during the process of digitization, stop the process
and, if possible, move the slide holder to a position where the slide can be easily removed by clicking
the
Emergency Manual Slide Removal Position
button (see section
4.3.6 Service tab
).
Warning!
Crush hazard! Before you remove a jammed slide, cancel
the running
digitization process, terminate the software session and turn the power
off.
Any manipulation performed on a mechanical component must be followed by a complete device
initialization routine (software restart).
Figure 74 – Removing the slide loader
March 13, 2012 - Rev. 1
3DHISTECH Ltd.
120(132)