Pannoramic SCAN 150 1.15 User’s Guide
5 Troubleshooting and Maintenance
HXP 120 illuminator
•
Keep inflammable and combustible materials outside the range of HXP 120 illuminator.
•
Do not expose yourself to the UV radiation of the HXP 120 illuminator.
•
When the HXP 120 illuminator is switched on:
◦
Never look directly into the fiber-optic cable.
◦
Never remove the fiber-optic cable from the device.
◦
Never connect the fiber-optic cable to the device.
◦
Never remove the housing from the device.
5.1.1 Cleaning the device
•
Before cleaning, switch off the device and pull the plug out.
•
Do not let liquid penetrate into the system.
•
To clean liquid from the device, use a lint-free cotton cloth. You can moisten the cloth with
water.
•
To clean greasy surfaces, it is recommended to apply
Isopropanol
or
Ethanol
on a piece of
lint-free cotton cloth and gently rub the surface.
5.1.2 Cleaning the magazine
Do not clean the magazines with xylene, because it might harm the plastic. We recommend to apply
Isopropanol
or
Ethanol
on a piece of lint-free cotton cloth and gently rub the surface instead. In case
the magazine is damaged, use either order a new one, as inserting a deteriorated magazine could
lead into unsafe loading of slides.
5.1.3 Preventive maintenance
To prevent fungus growth
•
Do not leave the equipment in an unconditioned environment with a storage temperature
above 60°C (140°F) or below 0°C (32°F), which may damage the equipment.
•
Maximum operating temperature is 35°C (95°F) with a relative humidity of less than 75%.
•
Install it in a room with proper air ventilation.
•
Do not let dust or fingerprints remain on optical surfaces.
March 13, 2012 - Rev. 1
3DHISTECH Ltd.
110(132)