Luna G3 BPAP 25A (G3 B25A) User Manual MID-97/v1.0
39 / 48
Problem
Possible Cause
Solution(s)
Water in mask
When the humidifier is used,
the humidified air tends to
condense in the cold
tubing and mask if the
room temperature is low.
Turn the humidity setting down
or raise the room temperature.
Place the tubing under the quilt
or use the tubing cover.
Hang the tubing loosely, the
lowest part of the tubing should
be lower than the patient's
head.
Nasal, sinus, or
ear pain
Sinus
or
middle
ear
inflammation
Contact
your
physician
immediately.
Discomfort due
to inability to
adapt to the
treatment
pressure
treatment pressure is a
prescription, and cannot
treat sleep apnea if the
treatment pressure is set
too low.
Patients may experience a
variance in time to acclimate
to their therapy.
If the problem persists, contact
your physician or home care
company for assistance.
Obstructive
sleep
apnea
symptoms recur
There are many causes of
recurring
symptoms
including weight change,
medication or alcohol, and
poor mask fit.
Contact your physician for
assistance.
The device is
too noisy
The
tubing
is
not
connected properly.
Reconnect the tubing properly.
Air
delivered
from the device
is
abnormally
hot
The air inlet of the device
may be partially blocked,
leading
to
insufficient
airflow into the device.
Replace the air filter (see 18.7
Replacing the Air Filter / PM2.5
Filter) and clean the air inlet.
Place the device in an area
where air flows freely, and
make sure the device is at least
20 centimeters away from the
wall, curtain, or other things.
Содержание G3 B25A
Страница 1: ...User Manual G3 B25A Luna G3 BPAP 25A ...