WARNING!
• Do not touch the heater plate of the device when it is working, otherwise you may
get burned. Turn off the heat when the heated humidifier is not in use.
CAUTIONS!
•
the humidity inside the mask is low when the water temperature is low.
• The greater the difference between the temperature inside the air tubing and
room temperature is, the more likely condensation can occur inside the tubing.
• If there are only a few condensed water droplets inside the tubing in the morning
after therapy, it means that the humidity level is appropriate; if there is lots of
condensed water droplets inside the tubing and / or mask, it means that the
humidity level is too high and should be set lower; Nasal dryness means that the
humidity level is too low and should be set higher.
15. Using the Cellular Module
The Luna G3 BPAP 25A with Cellular Module can wirelessly communicate with
iCodeConnect
®
. iCodeConnect
®
is intended to augment the standard follow-up
care of patients diagnosed with obstructive sleep apnea by displaying usage and
therapeutic information that has been transmitted from the patient's Luna G3 BPAP
25A therapy device to the clinician or healthcare professional.
(1) Insert the Cellular Module into the device and turn on the device. The device
screen displays the Main Screen shown in Fig. 15-1.
Fig. 15-1
(2) The Cellular Module starts searching for signals in a few seconds. Once a signal is
found, the module will automatically connect to it, and a signal icon will appear in
the status bar at the top of the device screen.
There are four different signal icons, as listed in Table 2:
Table 2 Description of Signal Icons