LM3000 has not been authorized for use on any product other than 3M model 1860 N95 respirators. 

 

Are there other measures to ensure the N95 respirator can be re-used? 

 

Re-use compatible N95 respirators only if they have been kept in a paper bag for a minimum of 5 days 

in addition to undergoing bioburden reduction using the Lumin LM3000. The compatible N95 respirator 

must be placed in the paper bag for a minimum of 5 days BEFORE use of the Lumin LM3000 in order to 

reduce contamination of the Lumin LM3000 when the respirator is placed in the drawer. 

 

Discard compatible N95 respirators after 4 bioburden reduction cycles or 5 donnings, whichever comes 

first. 

 

Caution: 

Compatible N95 respirators used in the Lumin LM3000 Bioburden Reduction UV 

System must be free of visible damage and soil/contamination (e.g., blood, respiratory or nasal 

secretions, or other bodily fluids; makeup; sunscreen; or other soil). Discard any soiled 

respirator. 

 

Caution: 

Sunscreens block UV light. The Lumin LM3000 may not be effective on N95 

respirators tainted with facially applied sunscreen. 

 

The instructions below provide some critical information regarding use/reuse of N95 respirators. For 

complete and up-to-date recommendations, please see CDC’s reuse recommendations, available at 

the website links provided above. 

 

 

Discard compatible N95 respirators after 4 bioburden reduction cycles or 5 donnings, whichever comes 

first.  

 

Discard any compatible N95 respirator whose traceability was lost (name or number of cycles are not 

clear). 

 

Discard compatible N95 respirators that were used during aerosol generating procedures.  

 

Discard compatible N95 respirators following contact with any patient co-infected with an infectious 

disease requiring respiratory or contact precautions.  

 

Discard compatible N95 respirators that are damaged, discolored, or visibly soiled.  

 

Discard any compatible N95 respirator that becomes hard to breathe through.  

 

Consider use of a cleanable face shield (preferred) over a compatible N95 respirator and/or other steps 

(e.g., masking patients, use of engineering controls), when feasible to reduce surface contamination of 

the respirator. 

 

Use a pair of clean gloves when donning a compatible N95 respirator and performing a user seal check.  

 

Discard gloves after the compatible N95 respirator is donned and any adjustments are made to ensure 

the respirator is sitting comfortably on your face with a good seal.  

 

Avoid touching the inside of the respirator. If inadvertent contact is made with the inside of the 

respirator, discard the respirator and gloves.  

 

Healthcare personnel must perform a user seal check of the compatible N95 respirator according to 

OSHA standard prior to beginning a shift. If the user seal check does not pass, discard the respirator. 

 

 

Below are some examples of engineering controls and practices for healthcare facilities to implement 

in addition to the use of the bioburden reduction system. For complete and up-to-date 

recommendations, please see CDC’s reuse recommendations, available at the website links provided 

above.  

 

 

Healthcare facilities shall perform a risk assessment to determine which additional CDC recommended 

Содержание Lumin LM3000

Страница 1: ...INSTRUCTIONS FOR HEALTHCARE FACILITIES LUMIN LM3000 BIOBURDEN REDUCTION UV SYSTEM USER S MANUAL MODEL LM3000 LM3000 ...

Страница 2: ...nless the authorization is terminated or revoked sooner Respirators that are NIOSH approved before bioburden reduction https wwwn cdc gov niosh cel only retain their NIOSH approval status post bioburden reduction if the respirator manufacturer permits the use of the bioburden reduction method with the specific system and cycle parameters To determine the NIOSH approval status of a specific bioburd...

Страница 3: ...3W Environmental Conditions Storage Temperature 0 to 120 F 18 to 49 C Operating Temperature 20 to 100 F 7 to 38 Humidity Up to 93 non condensing AC 110 V 60Hz AC 220 V 50Hz Voltage 12 25 x 8 5 x 7 75 Dimensions 5 5 Weight LM3000 SKU ...

Страница 4: ...5 respirator at a time by performing two 5 minute cycles one with the outer surface of the respirator facing up and one with the inner surface of the respirator facing up Two 5 minute cycles of 1 J cm2 i e one 5 minute cycle on each side of the compatible N95 respirator is considered one bioburden reduction cycle What are the known and potential benefits and risks of using bioburden reduced N95 re...

Страница 5: ...es for Optimizing the Supply of N95 Respirators during Shortages available at https www cdc gov coronavirus 2019 ncov hcp checklist n95 strategy html Consistent with CDC reuse recommendations it is critical that the respirator must be allowed to rest for 5 days to reduce the risk of inhaling pathogens that may remain in the filter material Bioburden reduction requires all surfaces and components o...

Страница 6: ...ost name or number of cycles are not clear Discard compatible N95 respirators that were used during aerosol generating procedures Discard compatible N95 respirators following contact with any patient co infected with an infectious disease requiring respiratory or contact precautions Discard compatible N95 respirators that are damaged discolored or visibly soiled Discard any compatible N95 respirat...

Страница 7: ... 5 days To minimize potential cross contamination store respirators so that they do not touch each other and the person using the respirator is clearly identified Storage containers must be disposed of or cleaned regularly Extend the use of N95 respirators by wearing the same N95 respirator for repeated close contact encounters with several different patients without removing the respirator per CD...

Страница 8: ... inspection of the compatible N95 respirator for the number of donnings or cycles completed correct operation of the Lumin LM3000 post cycle inspection of compatible N95 respirators for integrity and protocol to assure that the bioburden reduced compatible N95 respirator is returned to the original user The healthcare facility will routinely screen individuals performing bioburden reduction for CO...

Страница 9: ...te 5 minute cycle The Lumin LM3000 is pre programmed for a 5 minute cycle Caution Do NOT open the drawer before the cycle is completed If the drawer is opened before the cycle is complete the compatible N95 respirator has not been fully irradiated Repeat the complete 5 minute cycle 7 Change gloves 8 At conclusion of cycle confirm the UV indicator strip has changed color to confirm a full dose of U...

Страница 10: ... Schematic for one compatible N95 respirator Run a cycle with the respirator s outer blue surface facing up Remove the respirator maintaining the orientation and place on a clean paper towel Clean the base of the drawer Place respirator in drawer blue side down Discard the paper towel Protection of Bioburden Reduction Staff Select a location for handling of contaminated compatible N95 respirators ...

Страница 11: ...er 9 000 hours or 56 000 cycles In heavy use environments i e 10 uses daily it isrecommendedthatthe Lumin UVCbulbbe replaced annually to ensureoptimal efficacy Caution As the Lumin LM3000 Bioburden Reduction UV System relies on light reflection keep mirrored surfaces clean Use only water and a mild dishwashing liquid to carefully wipe mirrored surface Do not use any abrasive powders or liquids as ...

Страница 12: ...erse events of which they become aware related to the Lumin LM3000 and the bioburden reduced compatible N95 respirators This includes but is not limited to monitoring personnel using the Lumin LM3000 and healthcare personnel using the bioburden reduced compatible N95 respirators for signs and symptoms of potential infection with SARS CoV 2 or other respiratory infection and reporting such infectio...

Страница 13: ...in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation Please note that changes or modifications of this product is not expressly approved by t...

Страница 14: ...4 203AvenueANW Suite300 WinterHaven FL33881 www 3BLumin com and www 3Bproducts com Technical Support contact 3BLumin com support Sales and Information contact info 3BLumin com EPA Establishment No 93492 CHN 1 ...

Отзывы: