background image

10

 

 

14.

 

Change gloves.  

15.

 

Place a UV indicator strip in the drawer, facing up.  

16.

 

Close drawer and press button to initiate 5 minute cycle.  

17.

 

Change gloves.  

18.

 

At conclusion of cycle, confirm the UV indicator strip has changed color to confirm a full dose of UV has 

been delivered. 

 

Caution

: If the strip does not change color, the compatible N95 respirator has not been exposed to a 

sufficient dose of UV and must be put aside. Call 3B Medical for guidance and possible service. Do not 

use the device until 3B has provided instructions or serviced the device. 

 

19.

 

Remove the bioburden reduced, compatible N95 respirator from the Lumin LM3000.  

20.

 

Inspect the bioburden reduced, compatible N95 respirator for signs of damage to the respirator or 

elastic bands. If the respirator is damaged, the staff will discard the respirator and inform the user. 

21.

 

Add tick mark to the bioburden reduced, compatible N95 respirator.  

22.

 

Place the bioburden reduced, compatible N95 respirator in paper bag, label “READY FOR USE” and add 

user’s name.   

23.

 

Move the bioburden reduced, compatible N95 respirators to the area designated for “READY FOR USE” 

respirators. 

 

 

 

Schematic for one compatible N95 respirator: 

 

Run a cycle with the respirator’s outer (blue) surface facing up. 

 

Remove the respirator, maintaining the orientation, and place on a clean paper towel. 

 

Clean the base of the drawer. 

 

Place respirator in drawer blue side down. 

 

Discard the paper towel

 

Protection of Bioburden Reduction Staff 

•  Select a location for handling of contaminated, compatible N95 respirators and placement of the 

Lumin LM3000. It is recommended that the Lumin LM3000 be placed in a separate area, apart from 

any area used for clean/sterile supplies or food.  

•  Maintain the Chain of Custody Log Form at each bioburden reduction station. 

•  Handle all used, compatible N95 respirators as though they are contaminated. 

•  Don appropriate PPE (e.g., gloves, facial/respiratory protection, gown) when handling 

contaminated, compatible N95 respirators and while performing bioburden reduction to prevent 

transmission of pathogens between the respirators and bioburden reduction staff. 

•  Document the dates and names of all individuals whose compatible N95 respirators are placed in 

the Lumin LM3000 and the names of all persons performing the bioburden reduction. This will 

allow contact tracing in the event that an employee tests positive for COVID-19. 

•  Your facility will screen or test you regularly for COVID-19, as per facility practices. 

 

 

 

 

 

Содержание Lumin LM3000

Страница 1: ...INSTRUCTIONS FOR HEALTHCARE FACILITIES LUMIN LM3000 BIOBURDEN REDUCTION UV SYSTEM USER S MANUAL MODEL LM3000 LM3000 ...

Страница 2: ...nless the authorization is terminated or revoked sooner Respirators that are NIOSH approved before bioburden reduction https wwwn cdc gov niosh cel only retain their NIOSH approval status post bioburden reduction if the respirator manufacturer permits the use of the bioburden reduction method with the specific system and cycle parameters To determine the NIOSH approval status of a specific bioburd...

Страница 3: ...3W Environmental Conditions Storage Temperature 0 to 120 F 18 to 49 C Operating Temperature 20 to 100 F 7 to 38 Humidity Up to 93 non condensing AC 110 V 60Hz AC 220 V 50Hz Voltage 12 25 x 8 5 x 7 75 Dimensions 5 5 Weight LM3000 SKU ...

Страница 4: ...5 respirator at a time by performing two 5 minute cycles one with the outer surface of the respirator facing up and one with the inner surface of the respirator facing up Two 5 minute cycles of 1 J cm2 i e one 5 minute cycle on each side of the compatible N95 respirator is considered one bioburden reduction cycle What are the known and potential benefits and risks of using bioburden reduced N95 re...

Страница 5: ...es for Optimizing the Supply of N95 Respirators during Shortages available at https www cdc gov coronavirus 2019 ncov hcp checklist n95 strategy html Consistent with CDC reuse recommendations it is critical that the respirator must be allowed to rest for 5 days to reduce the risk of inhaling pathogens that may remain in the filter material Bioburden reduction requires all surfaces and components o...

Страница 6: ...ost name or number of cycles are not clear Discard compatible N95 respirators that were used during aerosol generating procedures Discard compatible N95 respirators following contact with any patient co infected with an infectious disease requiring respiratory or contact precautions Discard compatible N95 respirators that are damaged discolored or visibly soiled Discard any compatible N95 respirat...

Страница 7: ... 5 days To minimize potential cross contamination store respirators so that they do not touch each other and the person using the respirator is clearly identified Storage containers must be disposed of or cleaned regularly Extend the use of N95 respirators by wearing the same N95 respirator for repeated close contact encounters with several different patients without removing the respirator per CD...

Страница 8: ... inspection of the compatible N95 respirator for the number of donnings or cycles completed correct operation of the Lumin LM3000 post cycle inspection of compatible N95 respirators for integrity and protocol to assure that the bioburden reduced compatible N95 respirator is returned to the original user The healthcare facility will routinely screen individuals performing bioburden reduction for CO...

Страница 9: ...te 5 minute cycle The Lumin LM3000 is pre programmed for a 5 minute cycle Caution Do NOT open the drawer before the cycle is completed If the drawer is opened before the cycle is complete the compatible N95 respirator has not been fully irradiated Repeat the complete 5 minute cycle 7 Change gloves 8 At conclusion of cycle confirm the UV indicator strip has changed color to confirm a full dose of U...

Страница 10: ... Schematic for one compatible N95 respirator Run a cycle with the respirator s outer blue surface facing up Remove the respirator maintaining the orientation and place on a clean paper towel Clean the base of the drawer Place respirator in drawer blue side down Discard the paper towel Protection of Bioburden Reduction Staff Select a location for handling of contaminated compatible N95 respirators ...

Страница 11: ...er 9 000 hours or 56 000 cycles In heavy use environments i e 10 uses daily it isrecommendedthatthe Lumin UVCbulbbe replaced annually to ensureoptimal efficacy Caution As the Lumin LM3000 Bioburden Reduction UV System relies on light reflection keep mirrored surfaces clean Use only water and a mild dishwashing liquid to carefully wipe mirrored surface Do not use any abrasive powders or liquids as ...

Страница 12: ...erse events of which they become aware related to the Lumin LM3000 and the bioburden reduced compatible N95 respirators This includes but is not limited to monitoring personnel using the Lumin LM3000 and healthcare personnel using the bioburden reduced compatible N95 respirators for signs and symptoms of potential infection with SARS CoV 2 or other respiratory infection and reporting such infectio...

Страница 13: ...in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation Please note that changes or modifications of this product is not expressly approved by t...

Страница 14: ...4 203AvenueANW Suite300 WinterHaven FL33881 www 3BLumin com and www 3Bproducts com Technical Support contact 3BLumin com support Sales and Information contact info 3BLumin com EPA Establishment No 93492 CHN 1 ...

Отзывы: