Main Unit Weight
200g
Dimensions of device 158mm*56mm*51.5mm
Housing Materials of main
unit
ABS Plastic & Stainless Steel
Indicator
EMS Mode intensity Indicator
lights, Photon Mode working
indicator lights, Photon Mode
Indicator lights, Charging
Indicator
Operation Environment
Temperature: 5 ~ 40°C
Relative Humidity: ≤70% RH
(But not requiring a water vapour
Partial pressure greater than 50
hPa)
Barometric Pressure: 86~106
kPa
Temperature: -10 ~55°C
Relative Humidity: ≤80% RH
(But not requiring a water
vapour Partial pressure greater
than 50 hPa)
Barometric Pressure: 50~106
kPa
Storage & Transportation
Environment
EMS Mode Specification
Number of Output
channels
Output Intensity Level 3
1
Waveform and Shape
Pulsed Biphasic, Modulated
Square
Basic Pulse Duration 60ms
Pulse Frequency
8.33Hz
Maximum Output Voltage
(+/- 10%)
0.252V @ 500Ω
1.02V @ 2k
Maximum Output Current
(+/- 10%)
0.502mA @ 500Ω
0.510mA @ 2kΩ
Net Charge (per pulse) 42.96μC @500Ω
Maximum Phase Charge 33.264μC @ 500Ω
Maximum Current
Density
0.095mA/cm2@500Ω (The
Minimum Electrode Size:
2.91cm²)
Maximum Power Density
264.05μW/cm2@500Ω (The
Minimum Electrode Size:
2.91cm²)
Time Range
5 minutes
Photon Mode Specification
Wavelengths
Red RGB Light ( 620 - 730 nm
Wavelength ),
Green RGB Light ( 525 - 550 nm
Wavelength ),
Blue RGB Light ( 430 - 450 nm
Wavelength ),
White RGB Light (All the Red,
Blue and Green lights work
together).
Heat Temperature Range 41
℃
±2
℃
Energy
Red RGB Light (3.03X10-3
W.m-2.sr-1),
Green RGB Light
(0.106W.m-2.sr-1),
Blue RGB Light ( 4.1 W.m-2.
sr-1 ),
White RGB Light (4.7
W.m-2.sr-1).
Irradiance source
25 RGB lights
Time Range
5~7 minutes on each area
27
28
Содержание Time Keeper