Portable N

2

O Source Manual  Rev. B-2 

 

rotate  the  cup  a  seal  will  be  made.    You  should  rotate  the  cup  quickly  after 
puncturing  the  cartridge  to  minimize  loss  of  N

2

O  gas.    Do  not  tighten  the 

cartridge  cup  more  than  hand-tight  so  that  it  is  not  difficult  to  remove  the  cup 
later. 

 
 

Operation 

 

1)  When  not  using  the  Portable  N

2

O  Source,  keep  the  on/off  valve  closed  (knob 

turned completely clockwise). 
 

2)  To  use  the  Portable  N

2

O  Source,  check  that  there  is  tubing  connecting  the 

source  to  the  instrument  and  then  open  the  on/off  valve  by  turning  the  knob 
completely  counter-clockwise.    Use  only  moderate  hand  force  to  loosen  the 
knob (do not force with tools). 
 

3)  The  pressure  gauge  will  read  ~600  psi,  until  the  liquid  N

2

O  in  the  cartridge  is 

exhausted.  Then the pressure will rapidly fall off. 
 

4)  To  stop  using  the  N

2

O  source,  close  the  on/off  valve  by  turning  the  knob 

completely clockwise.  Do not tighten the knob more than hand-tight. 
 

 

Содержание 211

Страница 1: ...e N2O Source For use with the Model 211 Scrubberless Ozone Monitor and Model 408 NO Calibration Source Copyright 2017 2B Technologies Inc All rights reserved Technical Support www twobtech com techsup...

Страница 2: ...icological effects of N2O other than asphyxiation at extremely highly concentrations due to exclusion of oxygen N2O is not itself combustible but it strongly supports combustion of organic materials a...

Страница 3: ...to the back panel is a good idea to help ensure that the connections are remade correctly after installation of a new back panel Transfer all the components on the old back panel to the new one Mount...

Страница 4: ...from sliding Install the regulator mount on the left panel using the two 10 pan head screws and split locking washer supplied with the Portable N2O Source Install the left panel on the instrument and...

Страница 5: ...9 or email us at techsupport twobtech com Shipping Box Contents Open the shipping box and verify that it contains the following 1 Piercing device with integrated pressure regulator and removable inser...

Страница 6: ...ght 2 Close the on off valve by rotating the knob clockwise until you feel the knob stop turning 3 Unscrew the cartridge cup from the assembly Do not completely remove the cartridge cup until you are...

Страница 7: ...ob turned completely clockwise 2 To use the Portable N2O Source check that there is tubing connecting the source to the instrument and then open the on off valve by turning the knob completely counter...

Отзывы: