O Source Manual  Rev. B-2 

 

Description 

 
Nitrous  oxide  is  supplied  to  the  instrument  by  means  of  N

2

O  cartridges.    Each 

cartridge contains ~16 g or~ 8 g of liquid N

2

O. When using the Portable N

2

O Source 

with the Model 211 Scrubberless Ozone Monitor, the cartridges will supply a gas flow 
of N

2

O of 13 mL/min for ~8 hours (16-g cartridge) or 4 hours (8-g cartridge).  For the 

Model 408 NO Calibration Source, the cartridges will supply a gas flow of N

2

O of 40-

60 mL/min and provide an operating time of ~2 hours (16-g cartridge) or ~1 hour (8-g 
cartridge).    The  pressure  inside  the  cartridge  remains  nearly  constant  as  N

2

evaporates  and  flows  out  of  the  cartridge.    The  evaporation  of  N

2

O  causes  the  N

2

cartridge to cool, which, in turn, slightly reduces the internal pressure of ~600 psi. 
 

 

 
 

Installation of N

2

O Cartridges 

 

1)  Make  sure  that  any  cartridge  already  installed  in  the  assembly  is  empty  by 

checking that the reading on the pressure gauge is below 50 psi with the on/off 
valve  open.    To  make  sure  the  valve  is  open,  check  that  the  knob  cannot  be 
turned clockwise any further.  Do not tighten the knob more than hand-tight. 
 

2)  Close  the  on/off  valve  by  rotating  the  knob  clockwise  until  you  feel  the  knob 

stop turning. 
 

3)  Unscrew the cartridge  cup from the assembly.   Do not completely remove the 

cartridge  cup  until  you  are  certain  that  any  residual  N

2

O  gas  has escaped.    If 

you hear a hissing sound, N

2

O is escaping.  Once all the N

2

O has been vented, 

completely  remove  the  cartridge  cup  and  remove  the  empty  cartridge.    The 
cartridges are made of steel and may be recycled. 
 

4)  Insert  a  new  cartridge  into  the  cartridge  cup  with  the  narrow  neck  of  the 

cartridge pointed upward.   
 

If  using  an  8-g  N

2

O  cartridge,  the  removable  insert  should  be  left  in  place  in 

the  bottom  of  the  black  cartridge  holder.    If  using  a  16-g  cartridge,  slide  the 
insert out of the black cartridge holder. 

 

5)  Reattach the  cartridge  cup to the  piercing device.  As you screw the  cup onto 

the piercing device, the cartridge seal will be punctured, and as you continue to 

Warning:

    If  an  N

2

O  cartridge  is  punctured  outside  the  piercing 

device/regulator,  liquid  N

2

O  can  escape,  and  if  it  evaporates  on  skin  it  will 

cause  frost  bite.    The  cartridge  will  also  act  as  a  “rocket”  if  the  seal  is 
punctured  outside  the  piercing  device/regulator.    However,  the  N

2

cartridges  are  commonly  used  consumer  products  available  commercially 
and are safe to handle if the procedure below is followed.

 

Содержание 211

Страница 1: ...e N2O Source For use with the Model 211 Scrubberless Ozone Monitor and Model 408 NO Calibration Source Copyright 2017 2B Technologies Inc All rights reserved Technical Support www twobtech com techsup...

Страница 2: ...icological effects of N2O other than asphyxiation at extremely highly concentrations due to exclusion of oxygen N2O is not itself combustible but it strongly supports combustion of organic materials a...

Страница 3: ...to the back panel is a good idea to help ensure that the connections are remade correctly after installation of a new back panel Transfer all the components on the old back panel to the new one Mount...

Страница 4: ...from sliding Install the regulator mount on the left panel using the two 10 pan head screws and split locking washer supplied with the Portable N2O Source Install the left panel on the instrument and...

Страница 5: ...9 or email us at techsupport twobtech com Shipping Box Contents Open the shipping box and verify that it contains the following 1 Piercing device with integrated pressure regulator and removable inser...

Страница 6: ...ght 2 Close the on off valve by rotating the knob clockwise until you feel the knob stop turning 3 Unscrew the cartridge cup from the assembly Do not completely remove the cartridge cup until you are...

Страница 7: ...ob turned completely clockwise 2 To use the Portable N2O Source check that there is tubing connecting the source to the instrument and then open the on off valve by turning the knob completely counter...

Отзывы: