32
CG000542 | Rev A
User Guide | CytAssist Instrument | System Operation
Unbox
a.
Open box according to box label “this side up.”
Carefully cut or remove the tape on the top of the box and open the top flaps. Ensure
that cut is not too deep as it may damage the Instrument Accessory Kit box.
b.
Remove the Instrument Accessory Kit box, documents, and
packaging material.
c.
Lift the instrument from the sides. Place on bench top. Carefully remove
all packaging materials/tape from the exterior of the instrument.
Retain all packaging materials in the event that the instrument needs
to be moved (see
Moving & Shipping the Instrument
section).
d.
Remove Shipping Bolt from the back of the instrument. Retain for
future use (the Shipping Bolt holds the camera focus in a safe position
during transportation).
e.
Open lid. Carefully remove tape from the four Tissue Slide Clips and
the Visium Slide Lock as well as from the opposite end.
f.
Remove glass protective films from the alignment windows on both
stages. Close the lid (clicking sound indicates closure).
Step 1:
Installation
contd.
DO NOT POWER ON until ready for
Visium CytAssist
Accessory Kit
Lift instrument from sides
Retain packaging for future transportation
Remove shipping tape and then glass protective films
Visium Slide Stage
Tissue
Slide
Stage
Tissue Slide Clips
Tissue Alignment Windows
Visium Slide Lock
Visium Alignment Windows
Содержание Visium CytAssist
Страница 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications...
Страница 8: ...Safety Compliance Information 9 Safety 10 Regulatory...
Страница 11: ...System Components 12 System Components 13 Accessory Kits...
Страница 24: ...Menu Options 23 Menu Options 26 Firmware Update Procedure...
Страница 42: ...Maintenance 41 System Maintenance 43 Moving Shipping the Instrument...
Страница 48: ...Troubleshooting 47 Troubleshooting 49 Error Screens...