16
CG000542 | Rev A
User Guide | CytAssist Instrument | Tips & Best Practices
15 mm
align bottom edge
of frosted area* to the
horizontal black line
( 1 )
5 mm
5 mm
+ +
15 mm
( 2 )
5 mm
5 mm
If the slide has a frosted end or frosted marks along the bottom,
overlay on diagram 1 and then on diagram 2.
Due to variability in the dimensions of frosted areas across slide brands, the allowable
area for tissue placement is also variable across slide brands.
Check the allowable area from both the top (1) and bottom (2) to
ensure the tissue lies within the green allowable area:
•
15 mm
from edge of frosted area/marks
•
5 mm
from the sides
c.
Refer to the examples of good and bad tissue placement.
Refer to the Visium CytAssist Training Kit User Guide (CG000549) for further guidance.
If printing this page:
Select “actual size”
or “100%” to print
to scale. The slide
diagrams should each
measure 25 mm wide.
Examples of Good Tissue Placement
Blank Slide
++++++++
++++++++
Frosted Slide
Examples of Bad Tissue Placement
Blank Slide
++++++++
++++++++
++++++++
Frosted Slide
Check Tissue
Placement on Slide
contd.
align bottom edge** to
the horizontal black line
*If text is present below frosted area of slide,
align bottom of the letters to the line
**If markings are present at bottom edge
of slide, align markings to the
+
signs
Содержание 1000433
Страница 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications ...
Страница 8: ...Safety Compliance Information 9 Safety 10 Regulatory ...
Страница 11: ...System Components 12 System Components 13 Accessory Kits ...
Страница 24: ...Menu Options 23 Menu Options 26 Firmware Update Procedure ...
Страница 42: ...Maintenance 41 System Maintenance 43 Moving Shipping the Instrument ...
Страница 48: ...Troubleshooting 47 Troubleshooting 49 Error Screens ...