46
CG000542 | Rev A
User Guide | CytAssist Instrument | System Operation
c.
Open lid. Apply glass protective film or laboratory labeling tape to the
Visium and Tissue Alignment Windows (placement indicated by red
dashed line in the illustration below).
d.
Move the Visium Slide Lock to the closed position and tape down the
Visium Slide Lock and also the opposite end to secure the protective film.
e.
Move the four Tissue Clips to their innermost positions and tape down.
Ensure that the tape is placed over all four Tissue Slide Clips along
with the overlap with all four corners of the protective film.
f.
Close lid. Apply laboratory labeling tape to one side of the base, across
the top of the lid, and down the other side of the base (prevents the lid
from opening during transport).
Apply glass protective films and then shipping tape
Tissue
Slide
Stage
Visium
Slide
Stage
Tissue Alignment
Windows
Tissue Slide Clips
Visium Alignment Windows
Visium Slide Lock
Protective Film
Shipping Tape
Moving & Shipping
the Instrument
contd.
Содержание 1000433
Страница 5: ...Introduction 6 Introduction 7 Product Identification 7 Product Specifications ...
Страница 8: ...Safety Compliance Information 9 Safety 10 Regulatory ...
Страница 11: ...System Components 12 System Components 13 Accessory Kits ...
Страница 24: ...Menu Options 23 Menu Options 26 Firmware Update Procedure ...
Страница 42: ...Maintenance 41 System Maintenance 43 Moving Shipping the Instrument ...
Страница 48: ...Troubleshooting 47 Troubleshooting 49 Error Screens ...