User’s
Manual
13
analysing heart rhythm
low battery - replace battery
shockable rhythm detected
stay clear of patient
charging
press the shock button **
it is safe to touch the patient
begin cardiopulmonary resuscitation, now
make 5 cycles of 30 chest compressions followed by 2 breaths
press patient's chest down fast
(metronome – 30 beeps)
give two breaths
breath
(1sec mute)
breath
** if the shock button is not pressed within 10-12 seconds the unit will pass to CPR step along with the
voice message “
shock cancelled - shock button not pressed
”.
-------------------------------------------------------------------------------------------------------------------------------
S9 button
scenario #9
ADULT
Operations:
(i) shockable rhythm detection
(ii) faulty device warnings (red LED on)
Messages:
do not touch patient
analysing heart rhythm
shockable rhythm detected
stay clear of patient
charging
device failed - service required
-------------------------------------------------------------------------------------------------------------------------------
S10 button
scenario #10
ADULT
Operations:
(i) low battery warnings (red LED on)
(ii) shockable rhythm
(iii) faulty device warnings
Messages:
low battery - replace battery
do not touch patient
analysing heart rhythm
low battery - replace battery
shockable rhythm detected
stay clear of patient
charging
device failed - service required
-------------------------------------------------------------------------------------------------------------------------------
Extra 4 scenarios (from CS1 to CS4) are in option and could be customized under clients request.