Troubleshooting
4
Symptoms and Remedies
Consult the following table for help with troubleshooting the System. If the
problem you are experiencing is not listed in the table, contact your ZOLL service
representative for assistance.
Table 7-1. Symptoms, Probable Causes, and Corrective Actions.
Symptom
Probable Cause
Corrective Action
Red LED on control
panel is illuminated
and audible alarm
sounds.
System or control failure.
Discontinue use and contact ZOLL.
Red LED on control
panel is blinking and
audible alarm beeps
repeatedly.
Electrical power was
momentarily interrupted.
Turn off power and wait 30 seconds. Turn
on the System and restart. If problem
persists, discontinue use and contact ZOLL.
System will not start.
No electrical power.
Verify that the System is plugged into a
working circuit of correct capacity.
Assure that the power cord is securely
plugged into the power inlet connection of
the System.
The patient does not
cool. The activity
monitor indicates “
MAX
COOLING
” and the
patient’s temperature
is increasing.
The patient may be
febrile.
The fever has overcome the cooling cap-
ability of the catheter. Use additional patient
cooling methods as needed until the
patient’s temperature is stabilized.
The temperature control-
ler setting is incorrect.
Verify that the target temperature and that
the cooling rate option is not used.
The saline flow is
obstructed–indicated by
no rotation of the flow
indicator.
Inspect the entire length of tubing, from the
patient’s catheter connection to the System
and back to the patient. Clear all restrictions.
Check the flow indicator to confirm flow.
The saline flow rate is
incorrect.
Set the pump flow rate as specified by the
catheter’s
Instructions for Use
.
The catheter is impro-
perly positioned or is not
connected.
Place and position the catheter as directed
by the
Instructions for Use
.
Connect the saline circuit tubing to the
catheter and prime the saline circuit.
600479-001 Rev 2 128/147
Summary of Contents for thermogard xp
Page 1: ...Operation Manual 600479 001 Rev 2 1 147...
Page 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Page 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Page 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Page 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Page 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Page 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Page 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Page 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Page 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Page 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Page 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...