Accesories
-
HMIA
7
Figure 4
System Software Version
If your System does not have the compatible software version, then the correct
software will need to be installed for the HMIA to function. Failure to install the
correct software prevents the HMIA from functioning but it has no adverse affects
on the System.
Installation Equipment
The following items are necessary to install a HMIA.
1. System with correct software.
2. There are two different configurations of umbilical cables in the field. One
has a 6-pin connector and the other has an 8-pin connector. The
configuration of the HMIA corresponds to the type of connector on the
Umbilical Cable of the System being modified. HMIA part number
601033–001 is for the 8-pin circular connector and HMIA part number
601033–002 is for the 6-pin circular connector
3. TP-400 test fixture with adapter cable (Alsius Part number 500341-001)
that mates to the T1 connector of the System
4. Air trap assembly filled with water (Alsius Part number 500395-001).
5. The standard Hospital Monitor patient temperature cable that is used to
connect the Hospital Monitor to a Foley or rectal patient temperature
sensor.
6. Hospital Monitor that conforms to the YSI-400 requirements.
7. Optionally a Digital Multimeter may be used if a hospital monitor is not
available.
Installation Procedure
1. Disconnect power to the System
2. Disconnect the Umbilical Cable from the rear of the Display Housing
(See Figure 5)
600479-001 Rev 2 101/147
Summary of Contents for thermogard xp
Page 1: ...Operation Manual 600479 001 Rev 2 1 147...
Page 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Page 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Page 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Page 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Page 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Page 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Page 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Page 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Page 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Page 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Page 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...