Shipping and Storage Conditions
When shipping or storing the System, follow these recommendations:
Temperature range: -20ºC (-4ºF) to 60ºC (140ºF).
Atmospheric pressure range: 50 kPa to 106 kPa.
Do not allow liquids to spill on the System or its packaging.
Humidity range: 10% to 90% noncondensing.
Fragile contents, handle with care.
Always handle and store with the top facing up.
Ignition of Flammable Anesthetic Mixtures
The System is not category AP or APG equipment and must not be used in
environments where flammable anesthetic gas mixtures are present.
Electrical Hazards
This equipment has been tested and found to comply with the EMC limits of the
international standard
EN 60601-1-2
. These limits are designed to provide
reasonable protection against interference in a typical medical installation. The
equipment can radiate radio frequency energy if not installed in accordance with
the instructions, and may cause harmful interference to other devices in the
vicinity. There is no guarantee that interference will not occur in a particular
installation. Always comply with the following:
To avoid the risk of electrical shock, do not remove any panels of the
product.
Refer servicing to qualified personnel.
Never operate equipment with damaged power line cords.
Refer service and repairs to a qualified technician.
CAUTION! ELECTRIC SHOCK HAZARD.
Electric shock hazard. Always turn off the System and disconnect the
power line cord from the source before performing any service or
maintenance procedures, or before moving the System.
600479-001 Rev 2 9/147
Summary of Contents for thermogard xp
Page 1: ...Operation Manual 600479 001 Rev 2 1 147...
Page 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Page 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Page 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Page 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Page 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Page 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Page 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Page 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Page 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Page 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Page 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...