Operation
44
the flow indicator does not rotate freely during patient treatment, inspect the
entire tubing circuit for kinks or other restrictions to flow. Tapping on the flow
indicator will liberate any trapped air bubbles and return them to the saline bag.
Figure 4-53. Inline Flow Indicator.
WARNING! INVESTIGATE AIR TRAP ALARMS.
If an “Air Trap Fault” alarm occurs, it is likely that there is a leak in the tubing
circuit or the catheter has failed. Refer to
Chapter 6 – Alarms and Corrective
Actions
for assistance. Do not keep replenishing saline that is being rapidly
depleted–a problem exists that must be immediately remedied.
WARNING! START-UP KIT LIFETIME IS SEVEN (7) DAYS.
The designed operating lifetime for Start-up Kit components is seven (7) days
of continuous operation. If a patient must be treated for a longer period, a new
Start-up Kit must be installed in the System. Failure to adhere to this time limit
may cause injury to the patient.
600479-001 Rev 2 75/147
Summary of Contents for thermogard xp
Page 1: ...Operation Manual 600479 001 Rev 2 1 147...
Page 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Page 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Page 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Page 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Page 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Page 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Page 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Page 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Page 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Page 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Page 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...