Extended Storage

Do not store the 

SurePower Battery Pack

 for extended periods in a fully charged state. If you 

must store the 

SurePower Battery Pack

 for an extended period:

1. Charge the 

SurePower Battery Pack

 until three runtime LEDs light (this takes 

approximately 1 hour when charging a depleted battery).

2.

Every 4 months, recharge the 

SurePower Battery Pack

 until three runtime LEDs light.

Checking Capacity Loss

To check battery capacity (the amount of charge the battery can hold), press and hold down the 

Display

 button for 

at least

 5 seconds. 

Press and hold down the 

Display 

button for 

at least

 

5 seconds. 

While you hold down the 

Display

 button, the 

runtime indicator displays the battery runtime. 
When you release the 

Display

 button, the 

runtime indicator displays the battery capacity 
for 2 seconds, then redisplays the battery 
runtime.

During the 2 seconds after you release the 

Display 

button, 

the runtime indicator displays 

battery capacity and will light five LEDs to 
indicate that the capacity is equal or close to 
that of a new battery. 

Over time, as you use a 

SurePower Battery 

Pack

, the battery capacity decreases and this 

capacity check will light fewer LEDs. When the 
capacity check lights only a single LED, the 
battery is at or near 60% of its original capacity, 
and you should remove the battery from service.

If the 

Recalibration

 or 

Fault

 LED lights after 

you release the 

Display

 button, the runtime 

indicator will 

not

 display the capacity or run time 

for that battery.

Summary of Contents for SurePower

Page 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Page 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Page 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Page 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Page 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Page 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Page 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Page 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Page 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Page 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...