9650-0536-01 Rev. B
ZOLL
SurePower Battery Pack
Guide
5
Cleaning
Clean the
SurePower Battery Pack
with a soft damp cloth and mild soap.
Do not immerse or set the
SurePower Battery Pack
in fluids of any kind.
Do not
use alcohol or
ketones to clean the
SurePower Battery Pack
.
Maintenance
Perform a battery check, which verifies that the battery is fully charged and fault free, as part of
any regularly scheduled defibrillator test. Press the battery’s
Display
button to perform the
battery check.
As part of your defibrillator maintenance program, verify that the battery’s LEDs work (press
the
Display
button to test LEDs) -- if the LEDs do not work, remove the battery from service.
Also check the battery casing for cracks or other damage -- if the battery is cracked or damaged,
remove it from service.
SurePower Battery Pack
Icon
LED Indications
This is the
Recalibration
icon.
This LED lights for approximately 10 seconds (after you
press and release the
Display
button) if the battery needs to
be calibrated. If the
Recalibration
LED lights, the runtime
indicator will
not
display run time for that battery. For best
performance of the battery, you should recalibrate the
battery as soon as possible.
To manually recalibrate the
SurePower Battery Pack
, you
can insert the battery into the
SurePower
™
Charger Station
and perform a
ManualTest
(for more information, see the
ZOLL
SurePower Charger Station
Operator’s Guide
).
After you recalibrate the battery, the
Recalibration
LED will
only flash when you press the
Display
button.
This is the
Fault
Icon.
If the
SurePower Battery Pack
detects a battery fault, this
LED lights for approximately 10 seconds after you press the
Display
button.
If the
Fault
LED lights, the runtime indicator will
not
display
run time for that battery.
See the following section for more information on battery
faults.