background image

9650-0536-01 Rev. B

ZOLL

 SurePower Battery Pack

 Guide

5

Cleaning

Clean the 

SurePower Battery Pack

 with a soft damp cloth and mild soap. 

Do not immerse or set the 

SurePower Battery Pack 

in fluids of any kind. 

Do not

 use alcohol or 

ketones to clean the 

SurePower Battery Pack

.

Maintenance

Perform a battery check, which verifies that the battery is fully charged and fault free, as part of 
any regularly scheduled defibrillator test. Press the battery’s 

Display

 button to perform the 

battery check.

As part of your defibrillator maintenance program, verify that the battery’s LEDs work (press 
the 

Display

 button to test LEDs) -- if the LEDs do not work, remove the battery from service. 

Also check the battery casing for cracks or other damage -- if the battery is cracked or damaged, 
remove it from service.

SurePower Battery Pack

 

Icon

LED Indications

This is the 

Recalibration

 icon. 

This LED lights for approximately 10 seconds (after you 
press and release the 

Display

 button) if the battery needs to 

be calibrated. If the 

Recalibration

 LED lights, the runtime 

indicator will 

not

 display run time for that battery. For best 

performance of the battery, you should recalibrate the 
battery as soon as possible.

To manually recalibrate the 

SurePower Battery Pack

, you 

can insert the battery into the 

SurePower

 

 Charger Station 

and perform a 

ManualTest

 (for more information, see the 

ZOLL 

SurePower Charger Station 

Operator’s Guide

). 

After you recalibrate the battery, the 

Recalibration 

LED will 

only flash when you press the 

Display

 button.

This is the 

Fault

 Icon. 

If the 

SurePower Battery Pack

 detects a battery fault, this 

LED lights for approximately 10 seconds after you press the 

Display

 button.

If the 

Fault

 LED lights, the runtime indicator will 

not

 display 

run time for that battery.

See the following section for more information on battery 
faults.

Summary of Contents for SurePower

Page 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Page 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Page 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Page 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Page 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Page 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Page 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Page 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Page 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Page 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...

Reviews: