Using the ZOLL 

SurePower Battery Pack

The 

SurePower Battery Pack

 provides enhanced on-board processing that allows the battery to 

communicate its specific charging requirements to the ZOLL 

SurePower Charger Station

 or 

ZOLL defibrillator. The battery’s on-board processing also allows it to provide the following 
information:

The battery’s run time or capacity.

The battery’s calibration status.

Battery Faults Indications.

Figure 1-1 ZOLL 

SurePower Battery Pack

Note:

ZOLL ships the 

SurePower Battery Pack

 in a partially charged state and you must 

fully charge the battery before first use.

It is very important that you recharge a depleted 

SurePower Battery Pack

 as soon as possible.

Caution

Do not store fully depleted batteries or damage can result.

Caution

Do not store fully charged batteries for more than 90 days.

Summary of Contents for SurePower

Page 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Page 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Page 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Page 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Page 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Page 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Page 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Page 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Page 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Page 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...