background image

2

www.zoll.com

9650-0536-01 Rev. B

Do not use the 

SurePower Battery Pack

 in the presence of flammable agents (such as 

gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the 

presence of flammable agents can cause an explosion. 

Do not immerse or set the 

SurePower Battery Pack

 in liquid. Using the battery near or 

within puddles of water may present a shock hazard to the operator, patient, and nearby 

personnel.

Use only ZOLL chargers to charge ZOLL batteries.

If the display indicators on the 

SurePower Battery Pack

 are not consistent with the 

indicators on the 

SurePower Charger Station

 or ZOLL defibrillator, a fault condition may be 

present. Remove the battery from service as soon as practical, and verify the correct 

operation of the charger, defibrillator, and battery.

Cautions

Follow all recommended maintenance instructions. If a problem occurs, obtain service 

immediately. 

Do not use alcohol or ketones (MEK, acetone, etc.) to clean ZOLL rechargable defibrillator 

batteries. 

Do not sterilize or autoclave ZOLL batteries.

Avoid storage of fully depleted rechargable defibrillator batteries.

Do not install batteries into Monitor/Defibrillators or a 

SurePower Charger Station

 that is 

not plugged into live AC mains when storage may exceed 90 days. Battery damage may 

occur.

If you are disposing of a 

SurePower Battery Pack

, contact your local authorities to 

determine the requirements for the recycling and disposal of electrical equipment and follow 

those requirements.

Notes

ZOLL rechargable batteries may feel warm to the touch following a large number of shocks 

delivered at maximum energy or after recharging.

You can get additional information about the 

SurePower Battery Pack

 and its use from the 

related ZOLL charger and defibrillator manuals (for example, labeling information -- 

symbols used, Serial Number format, Date of Manufacture, and so on).

Summary of Contents for SurePower

Page 1: ...SurePower Battery Pack Guide 9650 0536 01 Rev B...

Page 2: ...t ZOLL Medical Corporation to determine if additional product information updates are available ZOLL is a registered trademark of ZOLL Medical Corporation SurePower is a trademark of ZOLL Medical Corp...

Page 3: ...fety hazard Should the SurePower Battery Pack be dropped inspect and test the unit prior to further use following the procedures described in this manual Avoid prolonged exposure of the SurePower Batt...

Page 4: ...enance instructions If a problem occurs obtain service immediately Do not use alcohol or ketones MEK acetone etc to clean ZOLL rechargable defibrillator batteries Do not sterilize or autoclave ZOLL ba...

Page 5: ...lows it to provide the following information The battery s run time or capacity The battery s calibration status Battery Faults Indications Figure 1 1 ZOLL SurePower Battery Pack Note ZOLL ships the S...

Page 6: ...proximately 10 seconds to indicate the battery s estimated run time If the battery is fully charged the LED over each of the five runtime indicator icons illuminates As the battery discharges the numb...

Page 7: ...ack Icon LED Indications This is the Recalibration icon This LED lights for approximately 10 seconds after you press and release the Display button if the battery needs to be calibrated If the Recalib...

Page 8: ...onds While you hold down the Display button the runtime indicator displays the battery runtime When you release the Display button the runtime indicator displays the battery capacity for 2 seconds the...

Page 9: ...ore the LED blinks The fault codes number from 1 to 29 To indicate fault code 1 the Fault LED blinks once To indicate fault code 2 the LED blinks twice and so on If the battery detects multiple batter...

Page 10: ...6 5 Length x 5 7 Width x 4 4 Height cm Weight 1 7 lb 77 kg Voltage 10 8 V Capacity 5 8 Ah Operating Temperature 32 F to 122 F 0 C to 50 C Optimal Charging Temperature 59 F to 95 F 15 C to 35 C Short T...