Inspection
9650-0535-01 Rev. C
SurePower Charger Station Operator’s Guide
4-1
Chapter 4
Maintenance and
Troubleshooting
This chapter describes a series of non-technical operational checks that you should perform
annually to ensure the proper operation of the
SurePower Charger Station
.
Inspection
Begin by inspecting the charger to ensure that it’s clean (with no fluid spills). Make sure that
nothing has been placed on top of the charger and that there are
at least
four inches of clear
space around the charger so that it can dissipate heat effectively.
You can then perform the following quick, non-technical operational checks:
1.
Inspect for cracks or damage to the charger housing.
2.
Inspect the AC cord for severe kinks or exposed wires.
3.
Connect the charger to live AC mains. The charger will beep, turn on the fan, and light all
LEDs for several seconds. The
Power
LED will remain lit after the other LEDs go off.
4.
With all battery bays empty, check that all LEDs on the control panel except the
Power
LED
are off.
5.
Press the
Test
button under each bay to verify that the bays are fully operational (for more
information, see the section, “Testing Battery Bays,” in Chapter 3).
6.
Insert a battery into each bay. Verify that the appropriate LEDs in each bay immediately
illuminate.
Cleaning
Clean the
SurePower Charger Station
with a soft cloth, mild soap, and water.
Do not immerse any part of the
SurePower Charger Station
in water. Do not use alcohol or
ketones (MEK, acetone, etc.). Do not autoclave the
SurePower Charger Station
.
Summary of Contents for SurePower