C
HAPTER
1 P
RODUCT
O
VERVIEW
1-4
www.zoll.com
9650-0535-01 Rev. C
Charging Methods
The
SurePower Charger Station
supports a variety of charging methods:
AutoTest
,
QuickCharge
, and
ManualTest
. These charging methods control how quickly the charger
brings the battery to full charge and whether or not the battery’s runtime indicator is
recalibrated.
AutoTest
The
AutoTest
charging method drains the remaining charge from a depleted battery before
recharging it. During this process, the charger measures the battery’s capacity to ensure that the
battery meets or exceeds the minimum runtime requirements for emergency defibrillator use.
The
AutoTest
charging method is the default charging method for ZOLL’s lead acid batteries
because it helps lead acid batteries maintain their storage capacity and extend their usable life.
QuickCharge
The
QuickCharge
charging method immediately charges the depleted battery.
The
QuickCharge
charging method is the default charging method for ZOLL’s lithium ion
batteries, such as the
SurePower Battery Pack
. The
QuickCharge
method is the quickest way
to charge lithium ion batteries, which do not require frequent testing to maintain their usable
life.
ManualTest
The
ManualTest
charging method tests and recalibrates ZOLL batteries.
ZOLL’s lead acid batteries with runtime indicators periodically require recalibration through
the use of the
ManualTest
charging method. The
SurePower Battery Pack
automatically
recalibrates itself.
We describe how to perform a
ManualTest
in Chapter 3.
Summary of Contents for SurePower
Page 1: ...SurePower Charger Station Operator s Guide 9650 0535 01 Rev C...
Page 22: ...CHAPTER 2 SET UP 2 4 www zoll com 9650 0535 01 Rev C...
Page 32: ...CHAPTER 3 OPERATION 3 10 www zoll com 9650 0535 01 Rev C...
Page 38: ...CHAPTER 5 USING ZOLL RECHARGABLE BATTERIES EFFECTIVELY 5 4 www zoll com 9650 0535 01 Rev C...