Indicator Light
When filtration has stopped the “Hold/Pause” diode is lit and the pH reading is done in the
normal way.
The “Alarm/Alarme” and “Hold/Pause” Diodes Flash.
Calibration error: the solution is not recognized. Press the “CAL” key to exit calibration and replace
the solution. If the problem persists, replace the sensor and recalibrate.
The “Alarm/Alarme” Diode Flashes and the “Hold/Pause” is Lit and the Pump has Stopped Dosing.
No flow in the circuit or the amount of corrective product is insufficient: to restore to normal
operation, restart filtration or change the bottle of product.
Please note: All alarm situations on the appliance are signalled by the red “Alarm/Alarme” diode flashing.
INDICATOR LIGHT ILLUMINATION