Instructions for use
3.1. General operation
3
7
Setting
target parameters
Press the "Rated parameter" button to open the
"Rated parameter" window.
Reminder
Entry field (1) gives the option of entering a specific number of Joules.
An acoustic signal sounds once the specified number of Joules has been
output. The signal repeats at the same intervals depending on the number of
Joules entered until the target total energy has been delivered.
If no target total energy is entered, an acoustic signal only sounds once the set
number of joules has been delivered.
Total energy
In the entry field (2) "Total energy", a target total energy can be set.
Once the target total energy is delivered, an acoustic signal sounds and the
laser application is stopped.
The application can be continued by pressing the footswitch again.
Input
reminder,
total energy
An "Input" window is opened by pressing the "Reminder" (1) and "Total
energy" (2) buttons.
Select the desired decimal position with the cursor or activate it with the touch
pen. The required values are entered with the numeric keypad.
Press "OK" to accept the new settings.
Press "Cancel" to cancel the process.
Note:
Setting a reminder and a target total energy is not necessary to carry out the
therapy.
(2)
(1)