Safety measures, commissioning,
function test
2.1 Quick operating instructions
2
3
Note:
Make sure that the main switch of the device is set to '0'.
Note:
Ensure that all persons present in the treatment room are wearing protective
goggles. Ensure that the applicator (6) is inserted completely into the
calibration sensor (8).
Installing the mains
cable
Connect the mains cable to the designated port (15) on the device and then
plug into the mains.
Interlock connector
Connect the interlock connector (27) to the appropriate socket on the device
(14).
Connecting the
footswitch
Plug the footswitch (24) into the appropriate socket (13) and then place it on
the floor.
Installing
spacers
Place the spacer on the applicator.
Switching on the
device
Switch on the device using the main switch.
Selecting a
programme
Activate the “Start” key and select a programme.
Activating the key
code
Activate the code key and enter the key code 1234.
Setting intensity
Set the required power.
Activating the laser
Activate the laser by pressing the “Start” key.
Applicator
Place the applicator in the correct position on the treatment area.
Starting treatment
Depress the footswitch to start the treatment.
Ending treatment
The treatment is interrupted or terminated by deactivating the footswitch.
After treatment, remove the spacer from the applicator and insert it into the
calibration sensor.