18
Warnings
8.
Cryo
Mini
is intended for operation by or on order of authorized medical
professionals only.
Should the Cryo
Mini
need to be turned onto its side for maintenance or
transportation, the device needs to stand upright for at least 30 minutes before it is
switched on to avoid compressor damage, and the defrosted water container
located at the side of the device needs to be checked to insure correct placement
Do not place the unit in direct contact with sources of heat (radiators, mud
preparations, saunas etc). Ensure a distance of at least 50 cm between the unit
and the wall to allow air inflow.
Unfavorable ambient conditions (e.g. ambient temperature above 86°F (30°C)
and high humidity) may lead to a reduction in performance (reduced cooling
capacity).
Magnetic and electric fields may affect the operation of the unit. Do not use
Cryo
Mini
close to equipment that generates strong electromagnetic fields (e.g. X-
ray or diathermy equipment or nuclear magnetic resonance tomography devices)
Dispose of packaging material properly. Keep such material out of reach of
children.
Cryo
Mini
is not designed for use in explosive or inflammable environments.
To avoid contaminating open wounds, the Cryo
Mini
can only be used before
injections (not during or afterwards).
To avoid freezer burns and over-cooling, the air stream should be directed evenly
over the entire area to be treated. Static cooling, or overly intensive cooling,
should be avoided.
Do not increase laser output power or energy when using Cryo
Mini
for skin
cooling. Always operate the laser in accordance with its user manual.
Skin cooling may reduce effectiveness of some laser treatments when laser has
the identical power output.
!
!
!
!