23
Safety
13.
Cryo
Mini
is manufactured in accordance to VDE 0750 part 1(DIN EN 60601-1)
safety regulations.
As the manufacturer, Zimmer MedizinSysteme GmbH assumes responsibility for
the safety and reliability of the unit only on the following conditions:
that the unit is supplied from an electrical socket with safety contact that meets
all regulations and the electricity supply conforms to DIN VDE 0100 part 710
that the unit is operated in accordance with the Instruction Manual
that any expansions, readjustments or modifications are carried out exclusively
by personnel acting on behalf of Zimmer MedizinSysteme GmbH.
The unit contains no user serviceable parts, with the exception of the Air Filter (7,
chapter 8.1, page 15) and the Defrosted Water Container (6, chapter 8.2, page 15).
Functional test
Cryo
Mini
performs an automatic self-test when switched on and checks the
operation of its various components.
If necessary, the operation of the cooling system may also additionally be checked
as follows:
1.
Switch on the unit.
2.
Wait until the Precooling Phase is complete and the start screen is
displayed. (see Adjustments, Settings and Menus).
3.
Press the Start/Stop button (15) to start operation.
4.
Select the various air flow levels and check the air flow rate and
cooling achieved.
Please observe any national regulations for recurrent check.