Manufacturer´s EMC declaration
18
Page 35
The field strength of stationary transmitters such as base stations for portable telephones and mobile land radios, wireless
keyboard stations, AM and FM radios and televisions cannot be accurately theoretically predetermined. To determine the
electromagnetic environment with regards to the stationary transmitters, a study of the electromagnetic phenomena of the
location should be considered. If the measured field strength at the location in which the cl
Trac
device will be used exceeds
the compliance level, the cl
Trac
device should be observed to verify proper functionality. If unusual performance
characteristics are observed, additional measures may be required, e.g. modified layout or a different location for the cl
Trac
device.
Interference resistance
tests
IEC 60601 test level
Compliance
level
Conducted disturbances
induced by RF fields
Radiated RF interference
variables in accordance
with IEC 61000-4-3
IEC 61000-4-6:2013
IEC 61000-4-3:2006
+A1:2007 +A2:2010
3 V , 150 KHz to 80 MHz
6 V , ISM bands 150 KHz to 80 MHz
6 V , Amateur radio bands 150 KHz to 80 MHz
3 V/m , 80 MHz to 2.7 GHz, AM 1khz 80%