
Operation Instructions
8
Page 15
(8)
Progression
•
Speed: Speed at which the force is built up. Expressed in force/second
(5N/s to 50N/s in steps of 1N/s)
(9)
Traction
•
Time: Total duration of this phase in minutes
(1min to 99min in steps of 1min)
•
Force: Continues force used in this phase
(15N to 900N in steps of 5N)
(10)
Regression
•
Speed: Speed at which the force is built down. Expressed in force/second
(5N/s to 50N/s in steps of 1N/s)
(11)
Notification bar
Shows the current selected program.
Notice:
As soon as the user enters a force of 200N or more, for safety reasons, a pop up shows
up and will ask to confirm this value.
Notice:
When no parameter is selected and the turning knob is used, the following error will be
displayed in the notification bar:
Notice:
The graphs shown on top of the screen during settings of the parameters are only meant
as an indicative view of the selection.
Activation patient
interrupt button
Once the correct parameters are installed and the patient is connected, press the patient
interrupt button once. It is strongly recommended to ask the patient to release the patient
interrupt button. This guarantees that the patient has handed over and is able to use the
patient interrupt button from that moment on.
Notice:
When the patient interrupt button is not pressed before the treatment, the following error
will be displayed in the notification bar:
(12)
Start button
After pressing the ‘Start’ button the
program parameter screen will be replaced by the
treatment screen.
Treatment screen
After selecting the (12) Start button the parameter setting screen will be replaced by the
treatment screen.