Cleaning
Disinfection
10
Page 21
Before starting any maintenance and cleaning measures the device must always be
switched off at the main switch and the mains cable unplugged.
Make sure that during cleaning and disinfection no liquids penetrate the device. Do
not use sprays.
If during cleaning or disinfecting liquid penetrates the device, please put the unit out
of service, protect it from being used again and contact your service representative.
Make sure that when cleaning and disinfecting the labels of the device (such as
warnings, labels of control devices, identification plate) are not damaged.
The device and its applied part are not considered critical in relation to hygiene when
used on non-injured and healthy skin.
Casing, Patient
interrupt button,
control knob and
swivel head
Tools:
-
Disposable wipes (cellulose, paper)
-
Alcohol-free plastic cleaner (e.g. cleaner for medical devices)
Cleaning (only manually)
In the event of visible contamination, the casing, patient interrupt button and all
cables can be cleaned using commercially available alcohol-free plastic cleaners.
Wipe the surface until the dirt is removed, using a soft cloth soaked according to the
specifications of the manufacturer of the cleaning agent but not dripping wet.
Disinfection (only manually)
Casing and patient button can be disinfected using disinfectant wipes. Use a
commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, veridical and fungicidal properties. Observe the application instructions
of the manufacturer. Wipe all surfaces using a cloth soaked according to the
specifications of the manufacturer of the disinfectant, but not dripping, or with cloth
pre-impregnated with disinfectant (wipes).
If applicable, also observe requirements for drying or post-cleaning.
Casing and patient button can be disinfected using disinfectant wipes. Use a
commercially available alcohol-free disinfectant for metal and plastic, with
bactericidal, veridical and fungicidal properties. Observe the application instructions
of the manufacturer. Wipe all surfaces using a cloth soaked according to the
specifications of the manufacturer of the disinfectant, but not dripping, or with cloth
pre-impregnated with disinfectant (wipes).
If applicable, also observe requirements for drying or post-cleaning.
Suitable Disinfection
agents
For manual disinfection, the following agents are suitable:
-
mikrozid® sensitiv wipes
-
Antifect FF
-
Gigasept FF
-
Quartamon Med
Observe the application information of the corresponding manufacturer!
Notice:
Only use the device in a completely hygienic environment.