Operation Instructions
8
Page 14
(5)
Traction
•
Time: Total duration of this phase in minutes
(1min to 99min in steps of 1min)
•
Max Force: Maximum force of this phase
(15N to 900N in steps of 5N)
•
Min Force: Minimum force of this stage
(10N to 895N in steps of 5N)
•
Max Hold Time: Hold time at maximum force in seconds
(1s to 60s in steps of 1s)
•
Min Hold Time: Hold Time at minimum force in seconds.
(1s to 60s in steps of 1s)
(6)
Regression
•
Speed: Speed at which the force is built down. Expressed in force/second
(5N/s to 50N/s in steps of 1N/s)
•
Steps: Number of steps
(1 step to 10 steps in steps of 1)
•
Hold Time: Time between two steps in seconds
(1s to 30s in steps of 1s)
Notice:
Parameter screen; Static
traction
Force can be displayed in Pound, Newton, or Kilogram. To change this, see chapter 8,
Settings.
Parameters can be changed before starting a treatment. To do this, select the related
parameter on the touch screen, and change the parameter by using the central control
knob.
(7)
Pre-Tension
•
Force: Force to which in this stage is built up
(10N to 900N in steps of 1N)
•
Time: Time for the pre-tension phase in seconds
(10s to 60s in steps of 1s)
7
8
9
10
11
12