Technical Information
9
Page 20
Power supply
110
–
240V
Mains fuse
2x F2.5A 250VAC
Protection class
IIb
Applied class
The patient interrupt button is a type B applied part.
Dimensions
390 (W) x 300 (H) x 350 (D)
Weight
Approx. 14 kg
Accuracy
Applied forces have an accuracy of 10% with respect to their set points
Storage/ Transport
Temperature-10
–
50°C
Humidity: 10
–
90% RH
Air Pressure: 700
–
1060 hPa
Operation
Temperature:10
–
25°C
Humidity: 20
–
80% RH
Air Pressure: 700
–
1060 hPa
Use environment
Indoor use only
Professional healthcare facility environment, except for near active HF
SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for
magnetic resonance imaging, where the intensity of EM DISTURBANCES is high.
Notice:
Storage and transport only in original packaging.
Subject to technical modifications.