background image

6

b)  The main TIME display reads “*60”, the default 

pressure set point, for approximately 1.5 seconds;

c)  The main TIME display increases (or decreases) 

by units of one until reaching a multiple of 5. Then 

the main TIME display increases (or decreases) by 

units of 5 to a maximum of 240 (or a minimum of 

0), thus altering the time set point;

d)  Release the touch-switch. This should cause the 

actual time of “0” to be displayed;

e)  Once again, press the main TIME INCREASE (or 

DECREASE) touch-switch; the main TIME display 

should read the last setting from step c (above);

f)  Release the main TIME touch-switch;

g)  Repeat steps

 a

 through 

above for the second 

TIME display functional check.

Notes for steps 3 & 4:

1.  Anytime an asterisk (*) is displayed in the left dis-

play digit, the data being displayed is its set point.

2.  The default factory set point for pressure is 250 mm 

Hg.

3.  The default factory set point for time is 60 minutes.

5.  Calibration Check

Note: 

During the power-up diagnostic self-test described 

above, the unit, which must hold a tolerance of ± 3 mm Hg, 

will test calibration. Should an out of calibration condition 

be detected, the unit will display “AMP” “FAIL” in the PRES-

SURE and TIME displays. Even though the unit completes 

this check during power-up, the following quantitative check 

is recommended at regular intervals.

a)  Set the unit to STANDBY by pressing and holding 

the ON/STANDBY 

(I/O)

 touch-switch. Note that the 

unit will not be set to STANDBY immediately. The 

ON/STANDBY touch-switch has a 2-second delay 

built in to help prevent accidental turn off.

b)  Wait approximately 10 seconds, then turn the unit 

back on by pressing the ON/STANDBY (

I/O

) touch-

switch. Press and hold the second PRESSURE 

(+) and (–) touch-switches until “CUFF” “CAL” is 

displayed.

c)  Connect a calibrated pressure meter, with a mini-

mum range of 0 to 700 mm Hg, to the Calibration 

Hose Assembly. See Fig. 1.

d)  Connect a pressure source capable of supplying 

(at minimum) 700 mm Hg of pressure. See Fig. 1.

e)  Insert the second cuff line with connector into the 

Second Cuff sense port. See Fig. 1.

f)  Insert the main cuff line with connector into the 

Main Cuff sense port. See Fig. 1.

g)  Apply 50 mm Hg of pressure to the cuff sense 

ports. The displays should each read 50 ±3 mm Hg.

h)  Increase the pressure to 250 mm Hg. Again, the 

displays should each read 250 ±3 mm Hg.

i)  Repeat step h (above) for 475 mm Hg.

j)  Remove the main and second cuff line from the cuff 

sense ports.

k)  Press and hold the second PRESSURE (+) and 

(–) touch-switches until “RES” is displayed in the 

second PRESSURE display. The main PRESSURE 

display should be reading 0 ±3 mm Hg.

l)  Insert the main cuff line with connector and second 

cuff line with connector into the reservoir sense 

ports. See Fig. 1.

m)  Apply 250 mm Hg of pressure to the reservoir 

sense ports. The display should read 250 ±3 mm 

Hg.

n)  Repeat step m for 475 mm Hg and 700 mm Hg.

o)  Remove pressure as well as the main and second 

cuff line with connectors from the unit.

p)  Set the unit to STANDBY by pressing and holding 

the ON/STANDBY (

I/O

) touch-switch.

NOTE: During the calibration check, if any reading is off 

by more than 3 mm Hg, the unit must be calibrated. See 

CALIBRATION in MAINTENANCE Section 3.

6.  Under Pressure Alarm Check

Create a leak in the main cuff while it is inflated. Make the 

leak large enough that even though the unit will be attempt-

ing to maintain pressure, the pressure still falls more than 

15 mm Hg below set point. Observe:

a)  The Main PRESSURE display window indicates 

low pressure by “LO-P”.

b)  The Alarm Silence Switch illuminates.

c)  After approximately 1 second, an audible tone will 

sound announcing the low pressure condition. This 

delay is used to preclude nuisance alarms.

d)  Stop the leak and observe that the pressure re-

turns to within limits, the audible tone ends, and the 

Alarm Silence Switch light extinguishes.

Repeat the above procedures for the second cuff.

2.3 INSTALLATION

The 

A.T.S. 

2000 Tourniquet is designed to be mounted on 

a table top or tourniquet stand (REF 60-4022-001). The 

adjustment range of the pole clamp on the right side of the 

unit will accommodate pole diameters of 0.5 to 1.5 inches 

(1.25 to 3.8 cm). Caution: Do not hang articles on the tour-

niquet stand which are not related to tourniquet use.
Inspect and ensure that the correct fuse drawer with ap-

propriately rated fuses is present. 100–120 V uses the grey 

fuse drawer with 1.0 A time delay fuses. 220–240 V uses 

the black fuse drawer with 500 mA time delay fuses.
Connect the power cord of the unit to a properly polarized 

and grounded power source whose voltage and frequency 

characteristics are compatible with those listed on the 

nameplate of the unit.
The 

A.T.S.

 2000 Tourniquet is now ready for use.

2.3.1 PRESSURE AND TIME DEFAULT 

SELECTIONS

An operator may want to modify the start up default pres-

sure of 250 mm Hg and the default time of 60 minutes for 

one or both cuffs.
The following steps will allow a user to customize the de-

fault settings for both cuffs.

1.  Prior to any cuff inflation, set the cuff target pres-

sure and the maximum inflation time to the new 

desired default values.

2.  Press the main cuff PRESSURE (+) and (–) touch-

switches simultaneously and hold until an audible 

tone is sounded, indicating that the new default 

values have been stored. The display should also 

read “UP” “DATE”.

The new pressure and time defaults will be written to non-

volatile storage and will provide the default pressure and 

Summary of Contents for A.T.S. 2000

Page 1: ...Operator Service Manual A T S 2000 TOURNIQUET SYSTEM REF 60 2000 101 00...

Page 2: ...EDIES In no case shall Zimmer Inc be liable for any special incidental or consequential damages whether based on breach of warranty or other legal theory Some states do not allow limitations on warran...

Page 3: ...ns 6 2 3 2 Volume Selection 7 2 4 Controls Indicators and Connectors 7 2 5 Single Cuff Operation 8 2 6 Dual Cuff Operation 9 2 7 Bier Block Cuff Operation IVRA 9 2 8 Alarm Conditions 9 2 8 1 Pressure...

Page 4: ...ssis From Rear Case 24 8 10 Separating Chassis 25 11 CPU Board 26 12 Power Supply Board 27 13 CPU Schematic Sheet 1 of 3 ENCLOSED 14 CPU Schematic Sheet 2 of 3 ENCLOSED 15 CPU Schematic Sheet 3 of 3 E...

Page 5: ...s time alarm low battery voltage hardware failure For certain types of equipment mal functions the unit will also display error messages for certain equipment malfunctions that identify the cause of e...

Page 6: ...Skin grafts in which all bleeding points must be readily distinguished Compromised vascular circulation e g peripheral artery disease Diabetes mellitus The presence of sickle cell disease is a relativ...

Page 7: ...ing effect exerted by an unpressurized cuff and its padding if used which prevents venous return at the beginning of the operation 2 By blood remaining in the limb because of insufficient exsangui nat...

Page 8: ...cuff line with connector and second cuff line with connector into the reservoir sense ports See Fig 1 m Apply 250 mm Hg of pressure to the reservoir sense ports The display should read 250 3 mm Hg n...

Page 9: ...etting touch switch or The TIME display will display the time setting with an asterisk in the far left digit The asterisk helps the operator identify that the data being displayed is the set point dat...

Page 10: ...uet can be fitted around the wrist Apply a leak free tourniquet cuff smoothly without wrinkles The valve port and hose connections should be placed so that the hose will not be kinked when the limb is...

Page 11: ...Table 2 1 The appropriate actions indicated are based on the most probable causes and should only be used as a guide Other causes of alarm conditions may indicate a need for other actions In addition...

Page 12: ...ESSURIZED DURING POWER UP This will occur if for example the unit is turned off and back on without deflating the cuff STEADY HIGH PITCH CUFF INFL LIT If a cuff pressure is 50 mm Hg or greater at the...

Page 13: ...OF CALIBRATION OR CAL VALVE MALFUNCTION CAL2 FAIL STEADY HIGH PITCH LIT SECOND CUFF OUT OF CALIBRATION OR CAL VALVE MALFUNCTION CALR FAIL STEADY HIGH PITCH LIT RESERVOIR OUT OF CALIBRATION OR CAL VALV...

Page 14: ...AC power or during patient transport The Windowing Watch Dog Circuit Safety Processor is a hardware circuit that monitors the timing signal generated by the microprocessor as well as valve states and...

Page 15: ...nd 475 13 Remove source pressure then disconnect the calibra tion connecting hose from both sense ports 14 Return the set points to zero by pressing the time touch switch 3 times for each cuff 15 Pres...

Page 16: ...aintenance is required of the battery charg ing circuit The life of the batteries depends on the type of service and the storage method Battery replacement will need to be more frequent with continued...

Page 17: ...nabled TP11 5 V 0 5 V 5 V Supply TP12 5 V 0 V 0 5 V 50 mV 5 V Nominal 0 V when depressing ON STANDBY switch on membrane panel TP13 V Common 50 mV Bulk Supply Common TP14 4 5 V 0 V 1 V 50 mV 0 V if Saf...

Page 18: ...ill not inflate a Touch Switch Panel connector not properly plugged into P9 b Tubing inside unit may be pinched or improperly connected c Valve s stuck d Pump not properly plugged into P10 2 Cuff s wi...

Page 19: ...ngs b Leak in valve manifold c Transducer not working TP29 Voltage will vary with pressure 0 25 to 4 V CPU Board 13 Valve fail alarm message a Faulty valve driver b No power to valves TP17 CPU Board T...

Page 20: ...07 Valve 24 V Burkert fast inflate deflate 60 2000 000 08 Fuse 500 mA 5 x 20 mm 62 1129 001 00 Contol Panel 13 touch switch 62 1131 001 00 Alarm Silence Switch without lamp 62 1132 001 00 Alarm Silen...

Page 21: ...cm 60 7500 005 30 in 76 cm x 4 125 in 10 5 cm 60 7500 006 34 in 86 cm x 4 125 in 10 5 cm 60 7500 007 42 in 107 cm x 4 125 in 10 5 cm A T S Contour Cuff Dual Port Single Bladder 60 7500 008 42 in 107...

Page 22: ...f 60 7400 002 Contour Arm Cuff 60 7400 003 Contour Thigh Cuff 60 7400 004 Lower Leg Cuff A T S Low Pressure Cuff Dual Port Dual Bladder 60 7400 010 Cylindrical Arm IVRA Cuff 60 7400 011 Contour Arm IV...

Page 23: ...y A Reservoir Sense Ports B Main Cuff Sense Port C Second Cuff Sense Port D Calibration Hose Assembly E Calibrated mmHg Pressure Meter minimum range of 0 to 700 mmHg F Pressure Regulator 0 to 700 mmHg...

Page 24: ...ON STANDBY I O Touch Switch B PRESSURE SETTING Touch Switches C TIME SETTING Touch Switches D INFLATE DEFLATE Touch Switches E ALARM SILENCE Switch F PRESSURE Displays G TIME Displays H AC Indicator L...

Page 25: ...23 3 Block Diagram...

Page 26: ...from Rear Case 1 Remove Pole Clamp knob Removing Chassis from Rear Case 2 Remove Rear Cover Screws Removing Chassis from Rear Case 3 Remove 4 Feet and Screws Removing Chassis from Rear Case 4 Slide F...

Page 27: ...25 8 9 10 Separating Chassis 1 Remove top 2 screws Separating Chassis 3 Access to all components Separating Chassis 2 Remove Recessed Bottom 3 screws...

Page 28: ...26 11 CPU Board...

Page 29: ...27 12 Power Supply Board...

Page 30: ...manual Electrical Hazard Year of manufacture Replace fuse as marked Conformity Marking of the Council of the European Community T V Product Service Munich Germany UL Classification Mark Medical Equipm...

Page 31: ...initions 17 WARNING A T S WILL NOT DEFLATE CUFF IN STANDBY MODE MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601 1 CAN CSA C22 2 NO 601 1 A...

Page 32: ...Revised 02 05 1996 2005 Zimmer Orthopaedic Surgical Products Printed in U S A 62 2000 101 00 Zimmer Orthopaedic Surgical Products 200 West Ohio Avenue P O Box 10 Dover Ohio U S A 44622...

Reviews: