11
Table 2.2 Hardware Malfunction Codes
PRESSURE
DISPLAY
TIME
DISPLAY
AUDIBLE TONE
ALARM
SILENCE
LIGHT
MEANING OF INDICATIONS
AMP
FAIL
STEADY HIGH
PITCH
LIT
AMPLIFIER OUT OF RANGE.
BAT LOW
PLUG IN
STEADY HIGH
PITCH
LIT
BATTERY VOLTAGE TOO LOW TO
ENSURE RELIABLE OPERATION.
BAT
FAIL
STEADY HIGH
PITCH
LIT
BATTERY DEFECTIVE OR TOO LOW TO
USE.
CALM
FAIL
STEADY HIGH
PITCH
LIT
MAIN CUFF OUT OF CALIBRATION OR
“CAL” VALVE MALFUNCTION.
CAL2
FAIL
STEADY HIGH
PITCH
LIT
SECOND CUFF OUT OF CALIBRATION
OR “CAL” VALVE MALFUNCTION.
CALR
FAIL
STEADY HIGH
PITCH
LIT
RESERVOIR OUT OF CALIBRATION OR
“CAL” VALVE MALFUNCTION.
MATH
FAIL
STEADY HIGH
PITCH
LIT
RESULT OF A MATH OPERATION WAS
OUT OF RANGE.
ROM
FAIL
STEADY HIGH
PITCH
LIT
MICROPROCESSOR FAILED A “ROM”
MEMORY CHECK.
RAM
FAIL
STEADY HIGH
PITCH
LIT
MICROPROCESSOR FAILED A “RAM”
MEMORY CHECK.
VALV
FAIL
STEADY HIGH
PITCH
LIT
AN IMPROPER VALVE COMBINATION
OCCURRED.
WDT
FAIL
STEADY HIGH
PITCH
LIT
WINDOWING WATCHDOG SYSTEM
DETECTED A MALFUNCTION.
Summary of Contents for A.T.S. 2000
Page 1: ...Operator Service Manual A T S 2000 TOURNIQUET SYSTEM REF 60 2000 101 00...
Page 25: ...23 3 Block Diagram...
Page 28: ...26 11 CPU Board...
Page 29: ...27 12 Power Supply Board...