30 |
Persona The Personalized Knee
Surgical Technique
NexGen Patella Peg Drill Guide
Sizes 29, 32, 35, 38, 41
00-5903-023-29/41
Patella Clamp
00-5903-020-00
NexGen 6.4 mm Patella/Femoral Drill
00-5120-052-01
2. Depress
1. Squeeze
Finish the Patella
(cont.)
Insert the appropriately-sized NexGen Patella Peg Drill
Guide into the patella clamp in the proper orientation
(Figure 65). Place the patella clamp with the NexGen
Patella Peg Drill Guide over the cut surface of the
patella, centered slightly toward the medial facet over
the 3.2 mm drill hole with the clamp oriented so two
of the holes are biased toward the medial side of the
patella (Figure 66).
Technique Tip:
For hard/sclerotic bone it may
be necessary to impact the NexGen Patella Peg
Drill Guide face to fully seat the drill guide. This is
necessary to assure that adequate bone removal is
attained during drilling to fully seat the implant.
Figure 66
Figure 68
Figure 67
Figure 65
Use the NexGen 6.4 mm Patella/Femoral Drill to drill
through the 3-peg holes in the NexGen Patella Peg
Drill Guide (Figure 67).
Technique Tip:
To facilitate unlocking the patella
clamp from the patella, apply slight gripping
pressure on the handles of the patella clamp and
depress the release lever to unlock the patella
clamp (Figure 68).