39 |
Compress Device
Surgical Technique
Spindle Implantation
(cont.)
Note:
If the compression cap cannot be removed
manually after one 180 degree turn, remove the nut
driver and place the spanner wrench (Figure 74) into
the surface holes of the compression cap and twist
off in a counterclockwise direction (Figure 75).
Important:
After the compression cap is removed
from the spindle,
DO NOT DISCARD.
Place the
compression cap in reserve on the back table until
the surgery is completed.
Figure 74
Figure 75