10 |
Compress Device
Surgical Technique
Anchor Plug
(cont.)
Slide the centering sleeve provisional (based on the
final reamer diameter B in the “Canal Preparation”
step) onto the anchor plug holder. Be sure to align the
exposed metal nub inside the provisional with the
female keyway located on the holder. When the
provisional meets the cross bar, lock into place with a
clockwise turning motion (Figures 13 and 14). Confirm
that the centering sleeve provisional fits snugly within
the canal at the osteotomy site.
Figure 13
Figure 14