29 |
Compress Device
Surgical Technique
Spindle
(cont.)
Place the spindle sizer over the exposed anchor plug
traction bar and select the spindle that comes closest
to one of the inscribed markings (Figure 52).
The selected implant must be closest to, but larger
than, the bone measured in this step (in this instance, a
38 mm spindle would be selected). The implant collar
of the spindle should overhang the bone (Figure 53).
Figure 52
Figure 53
A
B
C
D
Figure 54
Select a spindle based upon the prepared diameter
of the osteotomy site.
30 mm Spindle:
Anatomies up to 26 mm in diameter (Figure 54A).
38 mm Spindle:
Anatomies up to 34 mm in diameter (Figure 54B).
44 mm Spindle:
Anatomies up to 40 mm in diameter (Figure 54C).
49 mm x 37 mm Elliptical Spindle:
Anatomies in the metaphyseal flare where the bicortal
measurement does not exceed 45 mm (Figure 54D).