28 |
Compress Device
Surgical Technique
A
B
Transverse Pins
(cont.)
Remove the centering sleeve provisional from the
anchor plug holder (Figure 50). Set the provisional
aside, as it will be used later on in the technique.
Spindle
Diameter Selection
Measure the maximum diameter of the resected
bone with a ruler (Figure 51A) or the spindle sizer
(Figure 51B).
Note:
The ruler may be used to measure for all
four spindle sizes (30 mm, 38 mm, 44 mm and
37 x 49 mm elliptical). The spindle sizer measures
for the 38 mm spindle (small) and the 44 mm
spindle (large).
Figure 50
Figure 51