The reference figures can be found on pages 3 and 4 of this manual.
EN
Any modification that alters equipment
features from a safety and risk-prevention
point of view can only be carried out by the
Manufacturer who will then certify
conformity of the equipment with safety
standards. Therefore any modification or
maintenance intervention not considered in
this manual is to be considered prohibited.
Any modification, tampering or repair
performed by unauthorised staff along with
the use of non-original spare parts relieve
the Manufacturer from any liability. If the
necessary maintenance interventions are
not considered in this Use and Maintenance
Manual, contact the Authorised After-Sales
Centre.
8.4
RE-START AFTER LONG PERIOD OF INACTIVITY
If the equipment is re-started after a particularly
long period of inactivity, operate as indicated
in
Chapter
6
regarding
commissioning.
If the equipment functions anomalously, act as
described in Chapter 9 regarding “Diagnostics”. if the
problem should persist or is not considered, contact
the Authorised After-Sales Centre immediately.
8.5
hOW TO REQUEST AFTER-SALES ASSISTANCE
Choose one of the following options to request
after-sales assistance:
1. contact the Authorised After-Sales Centre;
2. telephone the Zhermack S.p.A. After-Sales
Service at +39 0425 597611 and proceed as
follows: give personal data (e.g. address and
telephone number); state the information given
on the equipment identification plate (see
paragraph 3.3); explain the problem with the
equipment clearly;
3. fax all of the information stated in the previous
point and the correctly filled-in identification
form (see Attachment 10.2) to the following n°
+39 0425 53596.
CHAPTER 9: DIAGNOSTICS
9.1
GENERAL WARNINGS
If any of the anomalies described below
should occur, solve the problem, where
allowed, with respect to all instructions
contained in this manual. If the problem
persists, contact the Authorised After-Sales
Centre.
9.2
TROUBLEShOOTING
MALFUNCTION The instrument placed in the
induction chamber does not heat
up.
CAUSE
1. The instrument is made from a
material which does not conduct
heat.
2. The instrument is tarnished (the
sensors detect light colours more
easily).
3. The sensors do not “detect” the
presence of the spatula.
4. The power cable is not connected.
5. There is no current to the socket.
6. The appliance is off.
7. Electrical fault.
REMEDY
1. Replace the instrument used.
2. Clean or replace the instrument used.
3. Adjust the sensitivity of the sensors.
4. Connect the power cable.
5. Check the socket/electrical panel in
the building.
6. Turn the appliance on by pressing
the ON/OFF button.
7. Contact an authorised service centre.
MALFUNCTION Red LED flashing rapidly.
CAUSE
1. An object has been left for too long
in the induction chamber.
2. The sensor is disorientated.
3. Electrical fault.
REMEDY
1. Remove the object from the
induction chamber.
2. Turn the tool heater off, then on
again. If the problem continues,
adjust the sensitivity of the sensors.
3. Contact an authorised service
centre.
MALFUNCTION The cooling fan makes a different
noise from usual.
CAUSE
1. Appliance overheated.
2. Faulty fan.
REMEDY
1. Turn the appliance off and allow it
to cool down.
2. Contact an authorised service
centre.