Why does my blood 

pressure fluctuate 

throughout the day?

Is the result the same 

if measuring on the 

right arm?

Why do I get a different 

blood pressure at home 

compared to the hospital?

1. Individual blood pressure varies 

multiple times everyday. It is also affected 

by the way you tie your cuff and your 

measurement position, so please take the 

measurement under the same conditions.

2.If the person takes medicine, the 

pressure will vary more.
3.Wait at least 3 minutes for another 

measurement.

The blood pressure is different even 

throughout the day due to weather, 

emotion, exercise etc, Also, there is the 

“white coat” effect, which means blood 

pressure usually increases in clinical 

settings.

It is ok for both arms, but there will 

be some different results for different 

people. We suggest you measure 

the same arm every time.

What you need to pay attention 

to when you measure your 

blood pressure at home: 

If the cuff is tied properly. 

If the cuff is too tight or too loose.

If the cuff is tied on the upper arm.

If you feel anxious.

Taking 2-3 deep breaths before 

beginning will be better for measuring.        

Advice: Relax yourself for 4-5 minutes 

until you calm down.

21

20

INFORMATION FOR USER

TROUBLESHOOTING

    This section includes a list of error messages and frequently 

asked questions for problems you may encounter with your blood 

pressure monitor. If the product is not operating as you think it 

should, check here before arranging for servicing.

PROBLEM SYMPTOM CHECK THIS

REMEDY

No power

Low

batteries

Error

message

Display
will not light up.

Batteries are exhausted. Replace with new batteries

Insert the batteries 

correctly

Replace with new batteries

Batteries are inserted

incorrectly.

Batteries are low.

EExx,shows on 

the display.

A calibration error

occurred.

Retake the measurement.

If the problem persists,

contact the retailer or our

customer service 

department for further 

assistance.Refer to the 

warranty for contact 

information and return 

instructions.

AC adaptor is inserted

incorrectly.

Insert the AC adaptor 

tightly

Display is dim or

shows

Warning

message

out shows

Out of measurement

range

the measurement result 

is out of the measurement 

range (SYS:60mmHg to 

230mmHg; or DIA:40mmHg

to 130mmHg;or Pulse: 

40-199 pulse/minute)

Data communication 

has failed

Make sure that phone’s 

Bluetooth is on or within 

the distance range

E 01 shows

E 03 shows

E 04 shows

The treatment of the

measurement failed.

E 02 shows

The cuff is too tight 

or too loose.

Refasten the cuff and then

measure again.

The monitor detected 

motion, talking,or the 

pulse is too poor 

while measuring.

Relax for a moment and 

then measure again. 

Relax for a moment and

then measure again.

The measurement 

process does not 

detect the pulse signal.

Loosen the clothing on

the arm and then 

measure again. 

E 12 shows

Summary of Contents for UAM-820BT

Page 1: ...品型号 材质 尺寸 70g书写纸 100 140 mm 单黑 产品名称 血压计 TMB 995 BS 004 李秋燕 2019 06 12 罗佳钰 2019 06 12 谭文昊 2019 06 12 TMB 995 BS 004 GB 08 PO 2F3502 TMB 995 BS 004 GB 08 说明书 A1 共 17 张 第 1 张 A 1 对应结构图纸 印色 1 1 表面处理 技术要求 1 黏合不可露胶 2 保持印刷面板上的清洁 3 注意套印的准确性 4 表面处理不可爆开 5 结构工艺以结构受控图为准 6 颜色参考 Manufactured for Zewa Inc 12960 Commerce Lakes Drive 29 Fort Myers FL 33913 USA www zewa com Toll Free Customer Service 1 888 993 3592...

Page 2: ... MANAGEMENT 15 Recall the Records Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 19 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbeat detector Why does my blood pressure fluctuate throughout the day Why do I get a different blood pressure at home compared to the hospital Is the...

Page 3: ... use only The Green Dot is the license symbol of a European network of industry funded systems for recycling the packaging materials of consumer goods Symbol for RECYCLE Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF APPLIED PARTS Symbol for DIRECT CURRENT Symbol for ENVIRONMENT PROTECTION Electrical waste products should not be disp...

Page 4: ...pairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060 2 2013 To verify the calibration of the AUTOMATED SPHYGMOMANOMETER pleas...

Page 5: ...e the soft cloth to clean the whole unit before and after use Don t use any abrasive or volatile cleaners The device doesn t need to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of ZEWA Don t open or repair the device by yourself in the event of malfunctions The device must ...

Page 6: ...s Replace the batteries whenever the below happens The shows The display dims The display does not light up BEFORE YOU START INTRODUCTION In order to get the best effect and protect your monitor please use the the right batteries and special power adapter which complies with local safety standard CAUTION CAUTION Do not use new and used batteries together Do not use different types of batteries tog...

Page 7: ...TION 1 2 Interference may occur in the vicinity of equipment marked with the following symbol And UAM 820BT may interfering vicinity electrical equipment Sensitive people including pregnant women pre eclamptic and those who implanted medical electronic instruments should avoid using the unit whenever possible Keep the monitor at least 20 centimeters away from the human body especially the head whe...

Page 8: ... on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measurement Wait at least 3 minutes between measurements This allows your blood circulation to recover The patient must relax as much as possible and do not move and talk during the measurement pro...

Page 9: ... measurement The Bluetooth symbol blinks on the LCD indicates data is transmitting 3 If the data is successfully transmitted the Bluetooth symbol will light up and the will disappear then the device will turn off If the data transmission fails the will disappear then the device will turn off 2 Hold and press button again to show the average value of the latest three records If the records are less...

Page 10: ...ath In a very cold environment Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Measurements may be inaccurate if taken in the following circumstances INFORMATION FOR USER The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so...

Page 11: ...pertension High Blood Pressure Hypertension Stage 1 High Blood Pressure Hypertension Stage 2 Hypertensive Crisis Consult your doctor immediately Systolic mmHg upper Diastolic mmHg lower less than 120 120 129 130 139 140 or higher Higher than 180 and or or and or less than 80 80 89 90 or higher Higher than 120 This chart reflects blood pressure categories defined by American Heart Association and l...

Page 12: ...lems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Batteries are exhausted Replace with new batteries Insert the batteries correctly Replace with new batteries Batteries are inserted incorrectly Batter...

Page 13: ...e 40 199 beat minute IP21 It means the device could protected against solid foreign objects of 12 5mm and greater and protect against vertically falling water drops Device Classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment A temperature range of 5 C to 40 C A relative humidity range of 15 to 90 non condensing but not requiring a water ...

Page 14: ...rt 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labelling and information to be supplied Part 1 General requirements IEC 80601 2 30 2009 A1 2013 Medical electrical equipment Part 2 30 Particular requirements for the basic safety and essential performance of automated non invasive sphygmomanometers ...

Page 15: ... domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines line s to line s 1 kV line s to earth 2 kV 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 300 cycle 30 A m 50Hz 60Hz NOTE UT is the a c mains voltage prior to application...

Page 16: ...ropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electrom...

Page 17: ... TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3 28 2 0 3 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 a For some services only the uplink ...

Reviews: