Why does my blood
pressure fluctuate
throughout the day?
Is the result the same
if measuring on the
right arm?
Why do I get a different
blood pressure at home
compared to the hospital?
1. Individual blood pressure varies
multiple times everyday. It is also affected
by the way you tie your cuff and your
measurement position, so please take the
measurement under the same conditions.
2.If the person takes medicine, the
pressure will vary more.
3.Wait at least 3 minutes for another
measurement.
The blood pressure is different even
throughout the day due to weather,
emotion, exercise etc, Also, there is the
“white coat” effect, which means blood
pressure usually increases in clinical
settings.
It is ok for both arms, but there will
be some different results for different
people. We suggest you measure
the same arm every time.
What you need to pay attention
to when you measure your
blood pressure at home:
If the cuff is tied properly.
If the cuff is too tight or too loose.
If the cuff is tied on the upper arm.
If you feel anxious.
Taking 2-3 deep breaths before
beginning will be better for measuring.
Advice: Relax yourself for 4-5 minutes
until you calm down.
21
20
INFORMATION FOR USER
TROUBLESHOOTING
This section includes a list of error messages and frequently
asked questions for problems you may encounter with your blood
pressure monitor. If the product is not operating as you think it
should, check here before arranging for servicing.
PROBLEM SYMPTOM CHECK THIS
REMEDY
No power
Low
batteries
Error
message
Display
will not light up.
Batteries are exhausted. Replace with new batteries
Insert the batteries
correctly
Replace with new batteries
Batteries are inserted
incorrectly.
Batteries are low.
EExx,shows on
the display.
A calibration error
occurred.
Retake the measurement.
If the problem persists,
contact the retailer or our
customer service
department for further
assistance.Refer to the
warranty for contact
information and return
instructions.
AC adaptor is inserted
incorrectly.
Insert the AC adaptor
tightly
Display is dim or
shows
Warning
message
out shows
Out of measurement
range
the measurement result
is out of the measurement
range (SYS:60mmHg to
230mmHg; or DIA:40mmHg
to 130mmHg;or Pulse:
40-199 pulse/minute)
Data communication
has failed
Make sure that phone’s
Bluetooth is on or within
the distance range
E 01 shows
E 03 shows
E 04 shows
The treatment of the
measurement failed.
E 02 shows
The cuff is too tight
or too loose.
Refasten the cuff and then
measure again.
The monitor detected
motion, talking,or the
pulse is too poor
while measuring.
Relax for a moment and
then measure again.
Relax for a moment and
then measure again.
The measurement
process does not
detect the pulse signal.
Loosen the clothing on
the arm and then
measure again.
E 12 shows