background image

Complied Standards List

EN ISO 14971:2012 / ISO 14971:2007 Medical devices - 

Application of risk management to medical devices

EN 1041:2008 Information supplied by the manufacturer of medical 

devices

EN 60601-1:2006+A1:2013/ IEC 60601-1:2005+A1:2012 Medical 

electrical equipment - Part 1: General requirements for basic safety 

and essential performance

EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical 

equipment - Part 1-11: General requirements for basic safety and 

essential performance - Collateral standard: Requirements for medical 

electrical equipment and medical electrical systems used in the home 

healthcare environment

EN 60601-1-2:2015/ IEC 60601-1-2:2014 Medical electrical 

equipment - Part 1-2: General requirements for basic safety and 

essential performance - Collateral standard: Electromagnetic 

disturbances - Requirements and tests

EN ISO 81060-1:2012 Non-invasive sphygmomanometers - Part 1: 

Requirements and test methods for non-automated measurement type

EN 1060-3:1997+A2:2009 Non-invasive sphygmomanometers - 

Part 3: Supplementary requirements for electro-mechanical blood 

pressure measuring systems

EN 1060-4:2004 Non-invasive sphygmomanometers - Part 4: Test 

procedures to determine the overall system accuracy of automated 

non-invasive sphygmomanometers

EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 

Medical electrical equipment - Part 1-6: General requirements for basic 

safety and essential performance - Collateral standard: Usability

IEC 62366-1:2015 Medical devices - Part 1: Application of 

usability engineering to medical devices

EN 62304:2006/AC: 2008 / IEC 62304: 2006+A1:2015   Medical 

device software - Software life-cycle processes

Risk management

Labeling
User manual

General Requirements 

for Safety

Electromagnetic

compatibility

Performance

requirements

Clinical investigation

Usability

Software life-cycle 

processes

Bio-compatibility

ISO 10993-1:2009 Biological evaluation of medical devices- Part 

1: Evaluation and testing within a risk management process

ISO 10993-5:2009 Biological evaluation of medical devices - 

Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological evaluation of medical devices - 

Part 10: Tests for irritation and skin sensitization

EN ISO 15223-1:2016 / ISO 15223-1:2016  Medical devices. 

Symbols to be used with medical device labels, labelling and 

information to be supplied. Part 1 : General requirements

IEC 80601-2-30:2009+A1:2013 Medical electrical equipment- Part 

2-30: Particular requirements for the basic safety and essential 

performance of automated non-invasive sphygmomanometers

ISO 81060-2:2013  Non-invasive sphygmomanometers - Part 2: 

Clinical validation of automated measurement type

25

24

This device complies with Part 15 of the FCC Rules. Operation is subject

to the following two conditions: (1) this device may not cause harmful

interference, and (2) this device must accept any interference received,

including interference that may cause undesired operation.

Caution: The user is cautioned that changes or modifications not

expressly approved by the party responsible for compliance could void

the user's authority to operate the equipment.

NOTE: This equipment has been tested and found to comply with the

limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.

These limits are designed to provide reasonable protection against

harmful interference in a residential installation. This equipment generates,

uses and can radiate radio frequency energy and, if not installed and used in

accordance with the instructions, may cause harmful interference to

radio communications. However, there is no guarantee that interference will

not occur in a particular installation.

If this equipment does cause harmful interference to radio or television

reception, which can be determined by turning the equipment off and on,

the user is encouraged to try to correct the interference by one or more

of the following measures:

-- Reorient or relocate the receiving antenna.

-- Increase the separation between the equipment and receiver.

-- Connect the equipment into an outlet on a circuit different from that to

which the receiver is connected.

-- Consult the dealer or an experienced radio/TV technician for help.

FCC Radiation Exposure Statement:

This equipment complies with FCC radiation exposure limits set forth for

an uncontrolled environment.

This transmitter must not be co-located or operating in conjunction with

any other antenna or transmitter.

FCC Statement

FCC ID:OU9TMB995-BS

COMPLIED STANDARDS LIST

EMC GUIDANCE

Summary of Contents for UAM-820BT

Page 1: ...品型号 材质 尺寸 70g书写纸 100 140 mm 单黑 产品名称 血压计 TMB 995 BS 004 李秋燕 2019 06 12 罗佳钰 2019 06 12 谭文昊 2019 06 12 TMB 995 BS 004 GB 08 PO 2F3502 TMB 995 BS 004 GB 08 说明书 A1 共 17 张 第 1 张 A 1 对应结构图纸 印色 1 1 表面处理 技术要求 1 黏合不可露胶 2 保持印刷面板上的清洁 3 注意套印的准确性 4 表面处理不可爆开 5 结构工艺以结构受控图为准 6 颜色参考 Manufactured for Zewa Inc 12960 Commerce Lakes Drive 29 Fort Myers FL 33913 USA www zewa com Toll Free Customer Service 1 888 993 3592...

Page 2: ... MANAGEMENT 15 Recall the Records Delete the Records INFORMATION FOR USER 17 Tips for measurement Maintenances ABOUT BLOOD PRESSURE 19 What are systolic pressure and diastolic pressure What is the standard blood pressure classification Irregular heartbeat detector Why does my blood pressure fluctuate throughout the day Why do I get a different blood pressure at home compared to the hospital Is the...

Page 3: ... use only The Green Dot is the license symbol of a European network of industry funded systems for recycling the packaging materials of consumer goods Symbol for RECYCLE Symbol for THE OPERATION GUIDE MUST BE READ Symbol for MANUFACTURER Symbol for SERIAL NUMBER Symbol for TYPE BF APPLIED PARTS Symbol for DIRECT CURRENT Symbol for ENVIRONMENT PROTECTION Electrical waste products should not be disp...

Page 4: ...pairment of patient blood circulation When measurement please avoid compression or restriction of the connection tubing The device cannot be used with HF surgical equipment at the same time The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER was clinically investigated according to the requirements of ISO 81060 2 2013 To verify the calibration of the AUTOMATED SPHYGMOMANOMETER pleas...

Page 5: ...e the soft cloth to clean the whole unit before and after use Don t use any abrasive or volatile cleaners The device doesn t need to be calibrated within two years of reliable service If you have any problems with this device such as setting up maintaining or using please contact the SERVICE PERSONNEL of ZEWA Don t open or repair the device by yourself in the event of malfunctions The device must ...

Page 6: ...s Replace the batteries whenever the below happens The shows The display dims The display does not light up BEFORE YOU START INTRODUCTION In order to get the best effect and protect your monitor please use the the right batteries and special power adapter which complies with local safety standard CAUTION CAUTION Do not use new and used batteries together Do not use different types of batteries tog...

Page 7: ...TION 1 2 Interference may occur in the vicinity of equipment marked with the following symbol And UAM 820BT may interfering vicinity electrical equipment Sensitive people including pregnant women pre eclamptic and those who implanted medical electronic instruments should avoid using the unit whenever possible Keep the monitor at least 20 centimeters away from the human body especially the head whe...

Page 8: ... on a flat surface Place your elbow on a table so that the cuff is at the same level as your heart Turn your palm upwards Sit upright in a chair and take 5 6 deep breaths Rest for 5 minutes before first measurement Wait at least 3 minutes between measurements This allows your blood circulation to recover The patient must relax as much as possible and do not move and talk during the measurement pro...

Page 9: ... measurement The Bluetooth symbol blinks on the LCD indicates data is transmitting 3 If the data is successfully transmitted the Bluetooth symbol will light up and the will disappear then the device will turn off If the data transmission fails the will disappear then the device will turn off 2 Hold and press button again to show the average value of the latest three records If the records are less...

Page 10: ...ath In a very cold environment Immediate measurement after tea coffee smoking When talking or moving your fingers When you want to discharge urine Measurements may be inaccurate if taken in the following circumstances INFORMATION FOR USER The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back one digit e g 2 becomes 3 and so...

Page 11: ...pertension High Blood Pressure Hypertension Stage 1 High Blood Pressure Hypertension Stage 2 Hypertensive Crisis Consult your doctor immediately Systolic mmHg upper Diastolic mmHg lower less than 120 120 129 130 139 140 or higher Higher than 180 and or or and or less than 80 80 89 90 or higher Higher than 120 This chart reflects blood pressure categories defined by American Heart Association and l...

Page 12: ...lems you may encounter with your blood pressure monitor If the product is not operating as you think it should check here before arranging for servicing PROBLEM SYMPTOM CHECK THIS REMEDY No power Low batteries Error message Display will not light up Batteries are exhausted Replace with new batteries Insert the batteries correctly Replace with new batteries Batteries are inserted incorrectly Batter...

Page 13: ...e 40 199 beat minute IP21 It means the device could protected against solid foreign objects of 12 5mm and greater and protect against vertically falling water drops Device Classification Battery Powered Mode Internally Powered ME Equipment AC Adaptor Powered Mode Class II ME Equipment A temperature range of 5 C to 40 C A relative humidity range of 15 to 90 non condensing but not requiring a water ...

Page 14: ...rt 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labelling and information to be supplied Part 1 General requirements IEC 80601 2 30 2009 A1 2013 Medical electrical equipment Part 2 30 Particular requirements for the basic safety and essential performance of automated non invasive sphygmomanometers ...

Page 15: ... domestic purposes Guidance and manufacturer s declaration electromagnetic immunity Immunity test 8 kV contact 15 kV air 8 kV contact 15 kV air 2 kV power supply lines line s to line s 1 kV line s to earth 2 kV 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle and 70 UT 25 30 cycles Single phase at 0 0 UT 300 cycle 30 A m 50Hz 60Hz NOTE UT is the a c mains voltage prior to application...

Page 16: ...ropriate for the frequency of the transmitter Recommended separation distances d 0 35 d 1 2 10V m 80 Am at 1kHz 80 MHz to 800 MHz d 1 2 800 MHz to 2 7 GHz d 2 3 where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electrom...

Page 17: ... TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulation b 18Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217Hz 2 0 3 28 2 0 3 28 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse modulation b 217 Hz 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 a For some services only the uplink ...

Reviews: