ZEUS-200S/150 ZO-OM-007
REV : 08 22 ZERONE
5.6
Using Method and Procedure
●
Monopolar
①
After confirming power and voltage (AC120V/AC230V), connect a ground power
cable to the power connection part on the back side of Electrosurgical Unit. If there
is no ground at Main Power, link an earth ground to the ground connection on the
back side of the product.
②
Turn on the Power Switch.
Check if Front Display indicates a number and then if REM alarm lamp flickers.
③
Connect Patient Return Plate Cable to the Patient Plate connection at the surgical
unit.
- Single Patient Plate
Connect Single Patient Plate to Patient Return plate Cable Clamp. Alarm lamp stops
flickering and the standby state starts. Attach Pad to a patient
’
s body.
- Double Patient Return Plate
Connect Double Patient Return Plate to Patient Return plate Cable Clamp. Attach Pad
to a patient
’
body. Alarm lamp stops flickering and the standby state starts.
④
Select Cutting Mode (Cut, Blend1, Blend2) and set a wanted output (W) by using
UP and DOWN buttons.
⑤
Select Coagulation Mode (Contact, Spray) and set a wanted output (W) level by
using UP and DOWN buttons.
⑥
In case of using Twin Button Handle, connect it to the connection 1 for Cutting
and Coagulation Electrode.
⑦
In case of using Foot Switch, connect Double Foot Switch Connector to Cutting,
Coagulation Foot Switch connection. If pressing Foot Switch, output is generated
from the Connection 2 for Cutting and Coagulation Electrode.
⑧
Press Twin Button Handle Switch or Double Foot Switch